Selection
The unit is rated for an operating temperature range of -20 to 200 °C, so it must be matched to the process temperature window.
Defines whether the bath can cover the required process temperatures before selection.
Source
“Operating temperature range -20...200 °C”
Performance
Cooling capacity drops with lower temperature, from 0.25 kW at 20 °C to 0.05 kW at -20 °C, so heat load must be checked at the coldest set point.
Prevents undersizing the refrigerating unit for low-temperature duty.
Source
“at 20°C 0,25 kW at 0°C 0,2 kW at -10°C 0,12 kW at -20°C 0,05 kW”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects room ventilation and ignition-source planning.
Flammable refrigerant classification drives installation safety and site requirements.
Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Installation
The power supply requirement is 110-115 V single-phase 50/60 Hz with a minimum 10 A fuse, so the electrical supply must be verified on site.
Mismatch with the available mains voltage or fuse rating prevents commissioning.
Source
“Power supply requirement 110-115V 1~ 50/60Hz”
Installation
The degree of protection is IP20, so the unit must be installed in a dry, non-condensing location protected from water ingress.
IP20 limits where the bath can be placed and how it can be cleaned.
Source
“Degree of Protection IP20”
Operation
Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may fall as ambient rises.
Room conditions directly affect achievable cooling performance and must be checked at the installation site.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic ordered.
Prevents chemical attack and pump failure when selecting the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.
Guides correct bath fluid preparation to avoid freezing damage.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Defines acceptable mains quality for reliable operation and warranty compliance.
Source
“-5% voltage and +2% frequency -> not allowed!”
Conformité
The equipment is EMC classified as Class A, Group 1 to EN55011, which has implications for use in residential environments.
Class A equipment may cause interference in domestic settings and affects site selection.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single/two-phase devices are delivered with a country-specific plug that must be specified at ordering.
Missing plug specification at order leads to unusable power connection on delivery.
Source
“with power cable and country-specific plug (please specify when ordering)”
Selection
An optional Pt100 sensor connection can display (not control) process temperature and is only available factory fitted.
Must be decided at purchase because it cannot be retrofitted in the field.
Source
“Option: Pt100 sensor connection #10688 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge).”