Minichiller 1000-H Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and key features of the Minichiller 1000-H
  2. Spec pp. 1-2
    Technical data, performance figures and operating limits
  3. Spec p. 2
    Detailed parameters according to DIN 12876 standard
  4. p. 2 table, 5 rows
  5. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Fuse16 A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S627670

Minichiller 1000-H — Technical data according to DIN 12876

ParameterValue
Order-No.3080.0017.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Cooling power at ambient temperature +20°C
at 15°C1 kW
at 0°C0,8 kW
at -10°C0,55 kW
at -20°C0,35 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Refrigerant quantity0,07 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Circulation pump:Pressure- and suction pump
max. delivery24 l/min
max. delivery pressure0,7 bar
max. delivery (suction)18 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
min. filling capacity2,8 l
Volume of expansion2,2 l
Overall dimensions WxDxH **280x511x454 mm
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current12,5 A

** Overall dimensions WxDxH — как напечатано в исходном тексте.

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit uses flammable refrigerant R-1270 (A3, H220), so the installation must respect flammable-refrigerant safety requirements.

A3 flammability affects room classification, ventilation, ignition-source control and local approval during selection and installation.

Source
“Refrigerant (ASHRAE, GHS) R-1270 (A3, H220)”
Installation

The power supply must be 208-240 V single/two-phase 50/60 Hz with a maximum current of 12.5 A and a 16 A fuse.

Defines the electrical feed, breaker/fuse rating and cabling needed before the chiller is installed.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not be combined in the unfavorable direction.

Prevents selecting a supply point whose combined voltage/frequency deviation would exceed the allowed operating envelope.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit is IP20, so it must be installed in a dry indoor location protected from water and dust ingress.

IP20 rules out washdown or outdoor installation and dictates enclosure/room protection.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop above 20 °C.

Room conditions directly limit available cooling capacity and must be checked against the process heat load.

Source
“max. ambient temperature 40 °C”
Performance

Cooling capacity falls from 1 kW at +15 °C to 0.35 kW at -20 °C, so the required setpoint determines the usable capacity.

Sizing must be based on capacity at the actual operating temperature, not the headline figure.

Source
“at -20°C 0,35 kW”
Selection

The plastic pump is not resistant to acetone or polyglycols, so these fluids require a different pump material.

Fluid compatibility must be confirmed before ordering; otherwise a more resistant plastic option is needed.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.

Prevents freezing and damage when operating near or below 0 °C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and the parts the customer must supply.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Conformité

The equipment is EMC Class A Group 1, which is intended for industrial environments and may cause interference in residential areas.

Class A limits where the unit may be installed without additional EMC measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The unit has adjustable overtemperature protection per DIN 12876 and safety classification III/FL.

The safety cut-out must be set correctly for the bath fluid and process to protect against overheating.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance

The refrigerant charge is only 0.07 kg with GWP 0, so refrigerant handling and leak reporting obligations are minimal.

Reduces regulatory burden and service logistics for the refrigeration circuit.

Source
“Refrigerant quantity 0,07 kg”