Sommaire

  1. Spec p. 2
    Technical data table according to DIN 12876 for the chiller
  2. p. 2 table, 9 rows

Technical data according to DIN 12876

CharacteristicValue
min. filling capacity2,8 l
Volume of expansion2,2 l
Overall dimensions WxDxH **280x490x424 mm
Power supply requirement208-240V 1~/2~ 50/60Hz
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/23

** Please respect space requirements. See operating conditions at www.huber-online.com

Engineering Facts

13 facts · every claim linked to a page
Selection

The unit is water-cooled and requires a cooling water differential pressure of at least 3 bar, so the site must provide a suitable chilled water supply.

Insufficient cooling water pressure will prevent the refrigeration machine from operating correctly.

Source
“min. cooling water differential pressure”
Installation

Cooling water supply pressure must not exceed 6 bar, so a pressure reducer may be needed on the building water line.

Exceeding the maximum pressure can damage the water-cooled refrigeration circuit.

Source
“max. cooling water pressure”
Installation

The unit is designed for cooling water intake between 5 °C and 20 °C, and pipes must be insulated if the water can drop below the dew point.

Out-of-range cooling water temperature reduces cooling power and can cause condensation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, so the specified 1 kW at 20 °C ambient is only valid at 15 °C cooling water inlet and 3 bar differential.

Sizing must account for actual site cooling water temperature, not just the nominal cooling capacity.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Safety

The refrigerant is R-290 (propane, A3 flammable), so the installation location must be assessed for flammable refrigerant safety.

A3 refrigerant requires appropriate safety measures and ventilation in the installation area.

Source
“R-290 (A3, H220)”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used in the circuit.

Using incompatible fluids will damage the pump and void reliable operation.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water below 0 °C.

Prevents freezing and damage when operating near or below 0 °C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.

IP20 is not water-resistant; outdoor or washdown environments are unsuitable.

Source
“Degree of Protection”
Installation

Ambient temperature must stay between 5 °C and 40 °C for correct operation.

Operating outside this range affects cooling performance and reliability.

Source
“min. ambient temperature”
Conformité

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Site power quality must be checked; some voltage/frequency combinations are explicitly prohibited.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The return line may reach 80 °C, so connected piping and components must be rated for this temperature.

Prevents damage to hoses and fittings exposed to hot return fluid.

Source
“Permissible temperature in return line 80 °C”
Maintenance

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water should be used.

Incompatible water chemistry can corrode or degrade circuit materials.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Selection

The pump delivers 24 l/min at 0.7 bar in normal mode but only 14 l/min at 0.2 bar in whisper mode, so flow requirements must match the chosen mode.

Whisper mode significantly reduces available flow and pressure for the process.

Source
“Delivery rate: 14 l/min”