Induztr
Résumé technique Visionneuse de documents Modèles associés

Minichiller 1200-H Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and main features of the Minichiller 1200-H
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 5 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Fuse16 A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S625033

Minichiller 1200-H — Technical data according to DIN 12876

ParameterValue
Order-No.3078.0016.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Cooling power at ambient temperature +20°C
at 15°C1,2 kW
at 0°C0,9 kW
at -10°C0,7 kW
at -20°C0,35 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Refrigerant quantity0,075 kg
CO2 equivalent0,0 t
Circulation pump:Pressure- and suction pump
max. delivery24 l/min
max. delivery pressure0,7 bar
max. delivery (suction)18 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
min. filling capacity2,8 l
Volume of expansion2,2 l
Overall dimensions WxDxH **280x511x454 mm
Net weight37 kg
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current12 A

** Overall dimensions WxDxH — как напечатано в исходном тексте.

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit is rated for -20 to 100 °C, so it suits applications within that range but not below -20 °C.

Defines the process temperature window the chiller can cover.

Source
“Operating temperature range -20...100 °C”
Selection

Cooling capacity drops with lower temperature, from 1.2 kW at 15 °C to 0.35 kW at -20 °C, so size the load against the actual operating point.

Prevents undersizing when the process runs cold.

Source
“at 15°C 1,2 kW at 0°C 0,9 kW at -10°C 0,7 kW at -20°C 0,35 kW”
Installation

The unit requires a 208-240 V single/two-phase 50/60 Hz supply with up to 12 A, so the electrical feed and fuse must be sized accordingly.

Determines the site power connection and protection rating.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz max. current 12 A”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed, so supply quality must be verified.

Avoids invalid supply conditions that could damage the unit or void compliance.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit is IP20 and must be operated between 5 and 40 °C ambient, so it needs a dry, climate-controlled location.

Constrains the installation environment and enclosure requirements.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump ordered.

Prevents chemical attack and premature pump failure.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.

Guides correct heat-transfer fluid preparation to avoid freezing damage.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Performance

Cooling capacity figures are valid at 20 °C room temperature and may drop if ambient rises, so derate for warmer rooms.

Affects capacity margin when the installation room is warm.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Single/two-phase devices come with a power cable and country-specific plug that must be specified at ordering.

Ensures the correct plug is supplied for the destination country.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Conformité

The equipment is EMC Class A Group 1, so it may cause radio interference in residential environments.

Relevant for installation in residential or sensitive EMC environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The unit has adjustable overtemperature protection per DIN 12876, so it must be set for the application's safe limit.

Critical safety setting to configure during commissioning.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance

The refrigerant is R-1270 (A3, H220) with GWP 0 and only 0.075 kg charge, so leak handling and flammability precautions apply.

Informs service safety and environmental handling requirements.

Source
“Refrigerant (ASHRAE, GHS) R-1270 (A3, H220)”