Induztr
Résumé technique Visionneuse de documents Modèles associés

Minichiller 1200w Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and main features of the Minichiller 1200w
  2. Spec p. 2
    Technical data according to DIN 12876 for the chiller
  3. p. 2 table, 15 rows
  4. p. 1 table, 24 rows

Technical data according to DIN 12876

ParameterValue
min. cooling water differential pressure3 bar
max. cooling water pressure6 bar
min. filling capacity2,8 l
Volume of expansion2,2 l
Overall dimensions WxDxH **280x490x424 mm
Net weight38 kg
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current5 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S571408

** Please respect space requirements. See operating conditions at www.huber-online.com

Minichiller 1200w — Technical data according to DIN 12876

ParameterValue
Order-No.3078.0007.01
Operating temperature range-20...40 (80)*** °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C at 15°C1,2 kW
Cooling power at ambient temperature +20°C at 0°C0,9 kW
Cooling power at ambient temperature +20°C at -10°C0,7 kW
Cooling power at ambient temperature +20°C at -20°C0,35 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,07 kg
CO2 equivalent0,0 t
Circulation pumpPressure- and suction pump
max. delivery24 l/min
max. delivery pressure0,7 bar
max. delivery (suction)18 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
Cooling water connectionG1/2 male

*** — сноска в исходном тексте не раскрыта. Дополнительно на странице указаны данные для whisper mode: switch for whisper mode; Reduced sound pressure level: 51 dB(A): +/- 4 dB(A); Pump data for whisper mode: Delivery rate: 14 l/min; Delivery pressure: 0.2 bar; Delivery rate (suction): 11 l/min; Delivery pressure (suction): 0.18 bar.

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water supply pressure between 3 bar minimum differential and 6 bar maximum, so the facility water system must be sized accordingly.

Insufficient or excessive cooling water pressure will prevent proper operation or damage the unit.

Source
“min. cooling water differential pressure 3 bar max. cooling water pressure 6 bar”
Installation

Cooling water inlet temperature must be kept between 5°C and 20°C, and pipes must be insulated if condensation occurs.

Out-of-range cooling water temperature reduces cooling capacity and causes condensation damage.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Selection

The unit uses flammable R-290 refrigerant (A3, H220), which affects safety classification and installation requirements.

Flammable refrigerant requires appropriate ventilation and safety measures at the installation site.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Selection

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump ordered.

Using incompatible media will damage the pump and void reliable operation.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Prevents freezing and damage in the cooling circuit at low temperatures.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Ensures the unit is not operated outside its electrical design envelope, avoiding malfunction.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit requires a minimum fuse of 10A and maximum of 16A on a 208-240V single/two-phase supply.

Correct fuse sizing protects the equipment and ensures compliance with electrical safety.

Source
“min. Fuse 10A max. Fuse 16A”
Operation

Whisper mode reduces sound pressure to 51 dB(A) but also reduces pump delivery to 14 l/min and 0.2 bar.

Engineers must account for reduced cooling performance when using quiet operation.

Source
“Reduced sound pressure level: 51 dB(A): +/- 4 dB(A)”
Installation

The unit has IP20 protection, so it must be installed in a dry, indoor environment protected from water ingress.

IP20 is not water-resistant; installation in wet or dusty areas is unsafe.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must be between 5°C and 40°C for proper operation.

Operating outside this range may cause overheating or control instability.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection

Cooling power drops significantly as temperature decreases, from 1.2 kW at +15°C to 0.35 kW at -20°C.

Selection must account for reduced capacity at low temperatures to avoid undersizing.

Source
“at 15°C 1,2 kW at 0°C 0,9 kW at -10°C 0,7 kW at -20°C 0,35 kW”
Installation

The return line permits up to 80°C, which must be considered when integrating with process loops.

Exceeding this temperature can damage internal components and seals.

Source
“Permissible temperature in return line 80 °C”