Minichiller 600

Sommaire

  1. Overview p. 1
    Product introduction and main features of Minichiller 600
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 3 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S595944

Minichiller 600 — Technical data according to DIN 12876

ParameterValue
Order-No.3066.0025.01
Operating temperature range-20...40 (80)*** °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 15°C0,6 kW
at 0°C0,5 kW
at -10°C0,35 kW
at -20°C0,15 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,085 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Circulation pump:Pressure- and suction pump
max. delivery24 l/min
max. delivery pressure0,7 bar
max. delivery (suction)18 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
min. filling capacity2,8 l
Volume of expansion2,2 l
Overall dimensions WxDxH **280x490x424 mm
Net weight37 kg
Power supply requirement110-120V 1~ 60Hz
max. current8,5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20

** Overall dimensions WxDxH; *** operating temperature range -20...40 (80) °C

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit's cooling capacity drops as ambient temperature rises, so size the chiller for the actual room temperature at the installation site.

Cooling capacity is rated at 20 °C ambient; higher ambient reduces available cooling and may require a larger model.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The chiller must be installed where ambient temperature stays between 5 °C and 40 °C.

Operating outside the ambient range risks poor performance or equipment damage.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The unit requires a 110–120 V single-phase 60 Hz supply with a 16 A fuse, so verify site power and breaker rating before installation.

Mismatched voltage/frequency or undersized fusing prevents safe commissioning.

Source
“Power supply requirement 110-120V 1~ 60Hz max. current 8,5 A Fuse 16 A”
Safety

The refrigerant is R-290 (propane, A3 flammable), so ignition sources and ventilation must be considered at the installation location.

A3 flammable refrigerant imposes site safety and ventilation requirements.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Safety

The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used as the heat transfer medium.

Using incompatible fluids will damage the pump and void reliable operation.

Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, use water mixed with glysantine or ethylene glycol rather than pure water.

Correct fluid mixture prevents freezing and protects the cooling circuit at sub-zero temperatures.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The pump connections are M16x1 male, and hose connection NW12 with M16x1 sleeve nuts are included, so plan matching plumbing.

Knowing the connection thread ensures correct hoses and fittings are ordered.

Source
“Pump connection M16x1 male”
Installation

The minimum filling capacity is 2.8 l and expansion volume is 2.2 l, so the external loop and fluid charge must be planned accordingly.

Underfilling or oversized external loops can cause cavitation or overflow.

Source
“min. filling capacity 2,8 l Volume of expansion 2,2 l”
Performance

The pressure-suction pump delivers up to 24 l/min at 0.7 bar pressure and 18 l/min at 0.4 bar suction, which bounds external circuit resistance.

External piping pressure drop must stay within these pump limits for adequate flow.

Source
“max. delivery 24 l/min max. delivery pressure 0,7 bar max. delivery (suction) 18 l/min max. delivery pressure (suction) 0,4 bar”
Operation

The permissible return-line temperature is 80 °C, which limits the heat load that can be returned to the chiller.

Exceeding return temperature can damage the unit or trip protection.

Source
“*** Permissible temperature in return line 80 °C”
Conformité

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Site power quality must meet these combined tolerances for warranty and safe operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.

IP20 is not water-resistant; outdoor or washdown locations are unsuitable.

Source
“Degree of Protection IP20”
Selection

The unit is classified as Class A, Group 1 for electromagnetic disturbance, which affects placement near sensitive equipment.

Class A equipment may require EMC precautions when installed near sensitive electronics.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are supplied with a country-specific plug that must be specified at ordering.

Failure to specify the plug type delays commissioning or requires field rework.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”