Product introduction and main features of the Minichiller 800w
Specpp. 1-2
Technical data, performance figures and operating limits per DIN 12876
p. 2table, 9 rows
p. 1table, 24 rows
Technical data according to DIN 12876
Parameter
Value
min. filling capacity
2,8 l
Volume of expansion
2,2 l
Overall dimensions WxDxH **
280x490x424 mm
Power supply requirement
208-240V 1~/2~ 50/60Hz
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/23
** Please respect space requirements. See operating conditions at www.huber-online.com
Minichiller 800w — Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-20...40 (80)*** °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at ambient temperature +20°C
at 15°C
0,8 kW
at 0°C
0,6 kW
at -10°C
0,45 kW
at -20°C
0,3 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Circulation pump
Pressure- and suction pump
max. delivery
24 l/min
max. delivery pressure
0,7 bar
max. delivery (suction)
18 l/min
max. delivery pressure (suction)
0,4 bar
Pump connection
M16x1 male
Cooling water connection
G1/2 male
min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar
*** см. сноску в оригинале; значения приведены дословно
Engineering Facts
12 facts · every claim linked to a page
Installation
The unit requires cooling water with a minimum differential pressure of 3 bar and a maximum pressure of 6 bar, so the site water supply must be verified before installation.
Insufficient cooling water differential pressure will prevent the water-cooled refrigeration machine from operating correctly.
Source
“min. cooling water differential pressure 3 bar”
Installation
The cooling water inlet temperature must be between 5°C and 20°C, and pipes must be insulated if the water temperature can fall below the condensation point.
Operating outside this cooling water temperature range will reduce cooling capacity and may cause condensation issues.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
The published cooling power is only valid at 20°C room temperature, 15°C cooling water inlet and 3 bar differential pressure, so actual capacity will differ at other conditions.
Engineers must not assume the catalog cooling power applies at their actual site conditions.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 3 bar differential pressure between cooling water inlet and -”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects safety classification and site risk assessment.
Flammable refrigerant classification requires appropriate safety measures and compliance checks during selection and installation.
Source
“R-290 (A3, H220)”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used; water with glysantine or ethylene glycol is recommended for freeze protection.
Using incompatible fluids will damage the pump and void reliable operation.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Conformité
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
The power supply must be checked against these combined tolerances to avoid non-compliant or unsafe operation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor environment protected from water ingress.
IP20 is not water-resistant, so installation location must avoid moisture and splashing.
Source
“Degree of Protection IP20”
Operation
The permissible return line temperature is 80°C, which defines the upper limit for the external circuit return flow.
Exceeding this return temperature can damage the unit or compromise safety.
Source
“Permissible temperature in return line 80 °C”
Installation
The minimum filling capacity is 2.8 l and the expansion volume is 2.2 l, which must be considered when sizing the external circuit and fluid volume.
Insufficient fluid volume or expansion space will cause poor temperature control and potential overflow.
Source
“min. filling capacity 2,8 l”
Selection
The pump delivers up to 24 l/min at 0.7 bar pressure and 18 l/min suction at 0.4 bar, which determines the achievable external circuit flow.
External circuit pressure drop must be matched to the pump curve to ensure adequate flow.
Source
“max. delivery 24 l/min”
Operation
Whisper mode reduces sound pressure to 51 dB(A) but also reduces pump delivery to 14 l/min at 0.2 bar, so quiet operation trades off hydraulic performance.
If low noise is required, the reduced pump performance must still satisfy the process cooling demand.
Source
“Delivery rate: 14 l/min”
Conformité
The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011, which may require EMC measures in sensitive environments.
Class A equipment may not be suitable for residential environments without additional EMC precautions.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”