Selection
The unit's operating temperature range is -100 to 40 °C, so it must be matched to the process temperature window.
Defines whether the cooler can cover the required process temperatures before selection.
Source
“Operating temperature range
-100...40 °C”
Performance
Cooling capacity drops sharply at the cold end (0.01 kW at -100 °C), so low-temperature duty needs careful capacity sizing.
Prevents oversizing assumptions and ensures the process heat load is removable at the coldest set point.
Source
“at -100°C
0,01 kW”
Installation
The unit requires a 110-120 V 1~ 60 Hz supply with 14 A max current and a 16 A fuse.
Determines site power provisioning, breaker/fuse rating and cable selection.
Source
“Power supply requirement
110-120V 1~ 60Hz
max. current
14 A
Fuse
16 A”
Installation
Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may fall if ambient rises.
Room conditions must be verified at the installation site to avoid derating or tripping.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
Voltage and frequency tolerances are coupled: ±5% voltage only with ±2% frequency, and -5% voltage with +2% frequency is not allowed.
Prevents installation on unstable or mismatched mains that could damage the compressor.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant is R-452A (A1, H280) with GWP 2141 and 0.26 kg charge, giving 0.6 t CO2 equivalent.
Affects refrigerant handling, leak response and F-gas/CO2-equivalent reporting obligations.
Source
“Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
0,26 kg
CO2 equivalent
0,6 t”
Safety
The second refrigerant stage is R-1150 (A3, H220), a flammable refrigerant, so ignition-source control is required.
A3 classification imposes flammable-refrigerant installation and ventilation precautions.
Source
“Refrigerant 2nd stage (ASHRAE, GHS)
R-1150 (A3, H220)”
Operation
The refrigeration unit runs continuously and temperature is controlled by a solenoid valve in the cooling circuit.
Explains control behaviour and why the compressor is not cycled on/off during normal operation.
Source
“The refrigeration unit works continuously. The temperature control actuates a solenoid valve in the cooling circuit.”
Installation
The probe is 48 mm in diameter and 155 mm long with a 1150 mm flexible connection, which constrains vessel port and routing.
Needed to confirm the immersion port size and the reach/routing of the cooling line.
Source
“Nominal diameter probe
48 mm
Length of probe
155 mm
Length flexible connection
1150 mm”
Conformité
The equipment is EMC Class A, Group 1, so it is intended for industrial, not residential, environments.
Class A equipment may cause interference in residential areas and must be sited accordingly.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at order.
Avoids a missing or wrong plug on delivery and delays at commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Conformité
The unit is compliant with US-SNAP for industrial process refrigeration and with applicable EU laws.
Confirms refrigerant acceptability for the intended end-use in regulated markets.
Source
“This equipment is compliant to US-SNAP and all applicable EU laws. The US-SNAP end-use for this equipment is the industrial process refrigeration.”