Technical specifications and performance data table
p. 1table, 29 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-45...100 °C
Cooling power
at 0°C
0,24 kW
at -20°C
0,18 kW
at -30°C
0,1 kW
at -40°C
0,05 kW
Safety classification
I / NFL
Refrigeration machine
air-cooled, natural
refrigerant
Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)
Global Warming Potential (GWP)
0,02
Refrigerant quantity
0,041 kg
CO2 equivalent
0,0 t
UN-number
UN 3358
Nominal diameter probe
13 mm
Length of probe
600 mm
Length flexible connection
1050 mm
Overall dimensions WxDxH **
190x295x360 mm
Net weight
16 kg
sound pressure level +/- 4 dB(A)
59 dB(A)
Power supply requirement
208-240V 1~/2~ 50/60Hz
max. current
1,4 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
S394802
** Please respect space requirements. See operating conditions at www.huber-online.com
Engineering Facts
12 facts · every claim linked to a page
Selection
The unit's cooling capacity is only 0.05 kW at -40 °C, so verify the required heat load at the lowest operating temperature before selecting it.
Cooling capacity drops sharply at low temperatures, which determines whether the immersion cooler can hold the process setpoint.
Source
“at -40°C
0,05 kW”
Safety
The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, so the installation area must be assessed for ignition sources and ventilation.
Flammable refrigerant classification affects site safety review, ventilation, and electrical zoning during installation.
Source
“R-290 (A3, H220)”
Installation
The device is rated IP20, so it must be installed in a dry indoor location protected from water and dust ingress.
IP20 limits where the unit can be placed and what washdown or outdoor conditions it can tolerate.
Source
“Degree of Protection
IP20”
Installation
Ambient temperature must stay between 5 °C and 40 °C, so the room climate must be checked before siting the unit.
Exceeding the ambient range risks reduced cooling capacity and equipment protection trips.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Performance
Cooling capacity figures are valid at 20 °C room temperature and may fall at higher ambient temperatures, so capacity margins must account for the actual room temperature.
Sizing on nominal capacity without ambient derating can lead to insufficient cooling in warm rooms.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Supply quality must be verified against this combined tolerance rule to avoid invalid operating conditions.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Single- and two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.
The plug type must be confirmed at order time to avoid rework at commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Conformité
The equipment is EMC Class A, Group 1, so it is intended for industrial environments and may cause interference in residential areas.
Class A classification affects EMC planning and suitability for non-industrial locations.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The minimum fuse is 10 A and maximum fuse is 16 A, so the upstream protection must be sized within this window.
Correct upstream fuse rating is required for safe electrical installation and warranty compliance.
Source
“min. Fuse
10A
max. Fuse
16A”
Selection
The probe is 13 mm in diameter and 600 mm long with a 1050 mm flexible connection, so vessel geometry and insertion depth must be checked.
Probe dimensions determine whether the cooler fits the vessel opening and reaches the required immersion depth.
Source
“Nominal diameter probe
13 mm
Length of probe
600 mm
Length flexible connection
1050 mm”
Operation
The refrigeration unit works continuously, so continuous operation and associated heat rejection to the room must be planned.
Continuous operation affects room heat load, energy consumption, and maintenance intervals.
Source
“The refrigeration unit works continiously.”
Conformité
The pressure equipment falls under Article 4.3 PED, so the applicable pressure equipment compliance route must be confirmed.
PED category determines documentation and conformity requirements for the installation.