Introduces the Unichiller 015w-H chiller and its main features
Specpp. 1-2
Lists performance, temperature, and dimensional data per DIN 12876
p. 2table, 11 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Volume of expansion
1,7 l
Overall dimensions WxDxH **
350x496x622 mm
Net weight
57 kg
sound pressure level +/- 4 dB(A)
54 dB(A)
Power supply requirement
220-240V 1~ 50Hz
max. current
13 A
Fuse
16 A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.1/24
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unichiller 015w-H (Order-No.: 3051.0010.01)
Characteristic
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Cooling power at ambient temperature +20°C at 15°C
1,5 kW
Cooling power at ambient temperature +20°C at 0°C
1 kW
Cooling power at ambient temperature +20°C at -10°C
0,7 kW
Cooling power at ambient temperature +20°C at -20°C
0,3 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,25 kg
CO2 equivalent
0,3 t
UN-number
UN 2857
Circulation pump
Immersion pump
max. delivery
29 l/min
max. delivery pressure
1 bar
Delivery at 0,5 bar
21 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
68 l/h
Consumption at water 15°C, flow 0°C
60 l/h
Consumption at water 15°C, flow -10°C
48 l/h
Consumption at water 15°C, flow -20°C
36 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,8 l
Cooling power values are given at ambient temperature +20°C for the respective flow temperatures (15°C, 0°C, -10°C, -20°C).
Engineering Facts
13 facts · every claim linked to a page
Installation
The chiller requires cooling water with a differential pressure of at least 1 bar and no more than 6 bar, so the facility water supply must be verified before installation.
Insufficient or excessive cooling water pressure will prevent proper operation or damage the unit.
Source
“min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar”
Installation
The unit is designed for cooling water intake between 5°C and 20°C, and if the water temperature falls below the condensation point the cooling water pipes must be insulated.
Operating outside this range or without insulation risks condensation, corrosion, and reduced performance.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase, so sizing must account for actual site water temperature.
Rated cooling power is based on 15°C water; higher site water temperatures reduce available capacity.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Selection
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic pump ordered at extra cost.
Using incompatible media will damage the pump and void reliable operation.
Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Prevents freezing and damage when operating at low temperatures.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.
Unsuitable water chemistry can corrode or foul these materials and impair heat transfer.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Conformité
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Ensures the unit is not operated outside its specified supply tolerances, which could cause malfunction.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The unit has IP20 protection and must be installed in an environment between 5°C and 40°C ambient temperature.
IP20 offers no water protection, and ambient limits affect cooling performance and safety.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Safety
The refrigerant R-449A has a GWP of 1397 and a charge of 0.25 kg, corresponding to 0.3 t CO2 equivalent, which matters for leak reporting and environmental compliance.
Determines regulatory obligations for refrigerant containment and leak checks.
Source
“Global Warming Potential (GWP)
1397
Refrigerant quantity
0,25 kg
CO2 equivalent
0,3 t”
Installation
The unit requires a 220-240V 1~ 50Hz supply with maximum current 13 A and a 16 A fuse.
Electrical supply and protection must be sized accordingly to avoid nuisance tripping or unsafe wiring.
Source
“Power supply requirement
220-240V 1~ 50Hz
max. current
13 A
Fuse
16 A”
Selection
The minimum filling capacity is 3.8 l and the expansion vessel volume is 1.7 l, which must be considered when selecting heat transfer fluid volume and expansion space.
Ensures sufficient fluid volume and proper expansion allowance for safe operation.
Source
“min. filling capacity
3,8 l”
Operation
The pump delivers 29 l/min maximum and 21 l/min at 0.5 bar, with a maximum delivery pressure of 1 bar, which constrains external circuit pressure drop.
External piping and equipment must be sized so the required flow is achievable within the pump curve.
Source
“max. delivery
29 l/min
max. delivery pressure
1 bar
Delivery at 0,5 bar
21 l/min”
Conformité
The equipment is classified for electromagnetic disturbance as EN55011 Class A, Group 1, which has implications for use in residential environments.
Class A equipment may cause radio interference and is not intended for residential use without mitigation.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”