Product introduction and model identification for Unichiller 017T
Specp. 2
Technical data according to DIN 12876 standard
p. 1table, 31 rows
Technical data according to DIN 12876 — Unichiller 017T (Order-No.: 3013.0067.01)
Parameter
Value
Operating temperature range
-10...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Temperature stability at -10°C
0,5 K
Safety classification
I / NFL
Cooling power at 15°C
1,7 kW
Cooling power at 0°C
0,9 kW
Cooling power at -10°C
0,4 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Circulation pump:
B
max. delivery
25 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
20 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 2,0 bar
10 l/min
Delivery at 3,0 bar
5 l/min
Pump connection
G3/4 male
min. filling capacity
2,5 l
Volume of expansion
2 l
Overall dimensions WxDxH **
450x510x1231 mm
sound pressure level +/- 4 dB(A)
64 dB(A)
Power supply requirement
230V 1~ 50Hz
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
1.0/21
** Overall dimensions WxDxH
Engineering Facts
12 facts · every claim linked to a page
Selection
The chiller is rated for an operating temperature range of -10 to 40 °C, so process requirements outside this window cannot be met.
Defines the process temperature envelope for which the unit can be selected.
Source
“Operating temperature range -10...40 °C”
Performance
Cooling capacity drops with falling temperature (1.7 kW at 15 °C, 0.9 kW at 0 °C, 0.4 kW at -10 °C), so the heat load must be checked at the lowest operating temperature.
Sizing must use the capacity at the actual set point, not the maximum rating.
Source
“Cooling power at 15°C 1,7 kW at 0°C 0,9 kW at -10°C 0,4 kW”
Performance
The circulation pump delivers 25 l/min max and 3 bar max pressure, with flow falling to 5 l/min at 3 bar, so system pressure drop must be matched to the pump curve.
Ensures the pump can overcome the external circuit resistance at the required flow.
The unit requires a 230 V single-phase 50 Hz supply, so the electrical installation must provide this supply and appropriate protection.
Determines the site power connection and cabling before installation.
Source
“Power supply requirement 230V 1~ 50Hz”
Installation
The chiller is IP20 and must be installed where ambient temperature stays between 5 and 40 °C, so it cannot be placed in a wet or hot environment.
Defines the permissible installation environment and enclosure protection limits.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation
The unit is air-cooled and needs space for airflow, so the stated space requirements around the housing must be respected.
Insufficient clearance will reduce cooling capacity and can cause overheating.
Source
“Please respect space requirements. See operating conditions”
Performance
Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises, so hot rooms reduce available capacity.
Sizing must account for the actual ambient temperature at the installation site.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Conformité
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
The site power quality must be checked against these combined tolerances before commissioning.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité
The equipment is EMC class A group 1, so it may cause radio interference in residential areas and is intended for industrial environments.
Affects installation location and EMC planning for sensitive equipment nearby.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.
The correct plug type must be ordered to match the site socket.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety
The refrigerant is R-449A with GHS classification H280 (pressurised gas) and GWP 1397, so refrigerant handling and leak reporting obligations apply.
Affects safety measures, ventilation and environmental compliance for the refrigerant charge.
Source
“R-449A (A1, H280)”
Operation
The unit is intended for externally closed applications with an atmospheric open expansion tank, so the external circuit must be closed and compatible with this design.
Prevents selecting the chiller for open-loop or unsuitable external circuits.