Unichiller 017T Technical Data

Sommaire

  1. Overview p. 1
    Product introduction and model identification for Unichiller 017T
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 31 rows

Technical data according to DIN 12876 — Unichiller 017T (Order-No.: 3013.0067.01)

ParameterValue
Operating temperature range-10...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,5 K
Safety classificationI / NFL
Cooling power at 15°C1,7 kW
Cooling power at 0°C0,9 kW
Cooling power at -10°C0,4 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pump:B
max. delivery25 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar20 l/min
Delivery at 1,0 bar17 l/min
Delivery at 2,0 bar10 l/min
Delivery at 3,0 bar5 l/min
Pump connectionG3/4 male
min. filling capacity2,5 l
Volume of expansion2 l
Overall dimensions WxDxH **450x510x1231 mm
sound pressure level +/- 4 dB(A)64 dB(A)
Power supply requirement230V 1~ 50Hz
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.1.0/21

** Overall dimensions WxDxH

Engineering Facts

12 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -10 to 40 °C, so process requirements outside this window cannot be met.

Defines the process temperature envelope for which the unit can be selected.

Source
“Operating temperature range -10...40 °C”
Performance

Cooling capacity drops with falling temperature (1.7 kW at 15 °C, 0.9 kW at 0 °C, 0.4 kW at -10 °C), so the heat load must be checked at the lowest operating temperature.

Sizing must use the capacity at the actual set point, not the maximum rating.

Source
“Cooling power at 15°C 1,7 kW at 0°C 0,9 kW at -10°C 0,4 kW”
Performance

The circulation pump delivers 25 l/min max and 3 bar max pressure, with flow falling to 5 l/min at 3 bar, so system pressure drop must be matched to the pump curve.

Ensures the pump can overcome the external circuit resistance at the required flow.

Source
“max. delivery 25 l/min max. delivery pressure 3 bar”
Installation

The unit requires a 230 V single-phase 50 Hz supply, so the electrical installation must provide this supply and appropriate protection.

Determines the site power connection and cabling before installation.

Source
“Power supply requirement 230V 1~ 50Hz”
Installation

The chiller is IP20 and must be installed where ambient temperature stays between 5 and 40 °C, so it cannot be placed in a wet or hot environment.

Defines the permissible installation environment and enclosure protection limits.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The unit is air-cooled and needs space for airflow, so the stated space requirements around the housing must be respected.

Insufficient clearance will reduce cooling capacity and can cause overheating.

Source
“Please respect space requirements. See operating conditions”
Performance

Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises, so hot rooms reduce available capacity.

Sizing must account for the actual ambient temperature at the installation site.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Conformité

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

The site power quality must be checked against these combined tolerances before commissioning.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité

The equipment is EMC class A group 1, so it may cause radio interference in residential areas and is intended for industrial environments.

Affects installation location and EMC planning for sensitive equipment nearby.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.

The correct plug type must be ordered to match the site socket.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The refrigerant is R-449A with GHS classification H280 (pressurised gas) and GWP 1397, so refrigerant handling and leak reporting obligations apply.

Affects safety measures, ventilation and environmental compliance for the refrigerant charge.

Source
“R-449A (A1, H280)”
Operation

The unit is intended for externally closed applications with an atmospheric open expansion tank, so the external circuit must be closed and compatible with this design.

Prevents selecting the chiller for open-loop or unsuitable external circuits.

Source
“For externally closed applications.”