Product introduction and main features of the chiller
Specpp. 1-2
Technical data, performance and operating parameters
Specp. 2
DIN 12876 compliant technical specifications table
p. 2table, 10 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Net weight
141 kg
sound pressure level +/- 4 dB(A)
60 dB(A)
Power supply requirement
230V 1~ 50Hz
max. current
16 A
Fuse
16 A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S338338
Technical data according to DIN 12876 — Unichiller 020Tw-H (Order-No.: 3024.0054.01)
Characteristic
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at 15°C
2 kW
Cooling power at 0°C
2 kW
Cooling power at -10°C
1,5 kW
Cooling power at -20°C
0,8 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,4 kg
CO2 equivalent
0,6 t
Circulation pump
B
max. delivery
25 l/min
max. delivery pressure
3 bar
Delivery at 0,33 bar
22 l/min
Delivery at 0,5 bar
20 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 2,0 bar
10 l/min
Delivery at 3,0 bar
5 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 0°C
90 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
2,5 l
Volume of expansion
4 l
Overall dimensions WxDxH **
400x440x1230 mm
** — сноска к габаритам (Overall dimensions WxDxH **) в тексте страницы не раскрыта.
Engineering Facts
11 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.
To ensure the cooling water circuit can deliver adequate flow and withstand the pressure without damage.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit is designed for cooling water intake temperatures between 5°C and 20°C, and pipes must be insulated if the water temperature falls below the condensation point.
Operating outside this range can reduce cooling capacity or cause condensation and damage.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.
To properly size the cooling water supply and predict performance under actual site conditions.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.
To avoid corrosion or damage by selecting compatible water quality and treatment.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations (e.g., -5% voltage with +2% frequency) are not allowed.
To ensure the power supply meets the allowed tolerances and avoid invalid combinations that could damage the unit.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Single/two-phase devices are supplied with a power cable and country-specific plug, which must be specified when ordering.
To ensure the correct plug type is supplied for the installation site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety
The refrigerant R-449A has a Global Warming Potential of 1397 and a CO2 equivalent of 0.6 t, which must be considered for environmental compliance.
To assess environmental impact and comply with refrigerant regulations.
Source
“Global Warming Potential (GWP) 1397”
Operation
The unit has a maximum ambient temperature of 40°C and a minimum of 5°C, limiting the installation location.
To ensure the chiller operates within its specified ambient conditions and maintains performance.
Source
“max. ambient temperature 40 °C”
Installation
The pump connection is G3/4 male and the cooling water connection is G1/2 male, requiring matching fittings.
To prepare correct mating connectors and hoses for installation.
Source
“Pump connection G3/4 male”
Maintenance
The minimum filling capacity is 2.5 l and the expansion volume is 4 l, which must be respected when filling the system.
To avoid insufficient filling or overflow during operation.
Source
“min. filling capacity 2,5 l”
Conformité
The equipment is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011.
To ensure electromagnetic compatibility with other equipment in the installation environment.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”