Unichiller 025Tw Technical Data

Sommaire

  1. Overview p. 1
    Introduces the Unichiller 025Tw model and its main features
  2. Spec pp. 1-2
    Lists performance, temperature, and dimensional specifications per DIN 12876
  3. p. 2 table, 8 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Power supply requirement208V 2~ 60Hz
max. current13 A
Fuse2x16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S542187

Technical data according to DIN 12876 — Unichiller 025Tw (Order-No.: 3054.0018.01)

ParameterValue
Operating temperature range-10...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at 15°C2,5 kW
Cooling power at 0°C1,2 kW
Cooling power at -10°C0,6 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,4 kg
CO2 equivalent0,6 t
Circulation pumpB
max. delivery30 l/min
max. delivery pressure3,8 bar
Delivery at 0,12 bar29 l/min
Delivery at 0,5 bar25 l/min
Delivery at 1,0 bar21 l/min
Delivery at 1,5 bar17 l/min
Delivery at 2,0 bar13,5 l/min
Delivery at 2,5 bar10 l/min
Delivery at 3,0 bar6,5 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C120 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity2,5 l
Volume of expansion4 l
Overall dimensions WxDxH **400x440x1230 mm
Net weight138 kg

** см. сноску в оригинале (Overall dimensions WxDxH **)

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure of at least 1 bar and no more than 6 bar at the inlet.

Ensures the water supply system can provide adequate flow without exceeding pressure limits.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is designed for cooling water intake temperatures between 5°C and 20°C, and pipes must be insulated if below the condensation point.

Prevents condensation and ensures specified cooling performance.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, and flow rate may need to increase.

Affects sizing of the cooling water supply and expected performance.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so only compatible water quality should be used.

Prevents corrosion and damage to the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité

Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but combinations like -5% voltage and +2% frequency are not allowed.

Ensures the power supply meets the unit's electrical requirements.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.

Avoids delays and ensures correct plug type for the installation site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The refrigerant R-449A has a Global Warming Potential of 1397 and a CO2 equivalent of 0.6 t, which must be considered for environmental regulations.

Affects compliance with refrigerant regulations and leak reporting requirements.

Source
“Global Warming Potential (GWP) 1397”
Operation

The unit has a safety classification I / NFL according to DIN 12876.

Indicates the safety level for use with flammable or non-flammable fluids.

Source
“Safety classification I / NFL”
Installation

The minimum filling capacity is 2.5 liters and the expansion volume is 4 liters.

Ensures proper fluid volume for safe operation and expansion.

Source
“min. filling capacity 2,5 l”
Installation

The pump connection is G3/4 male and the cooling water connection is G1/2 male.

Needed to select matching fittings and hoses for installation.

Source
“Pump connection G3/4 male”
Operation

The unit is for externally closed applications only.

Prevents misuse in open systems that could cause contamination or overflow.

Source
“For externally closed applications.”
Installation

The maximum ambient temperature is 40°C and minimum is 5°C.

Ensures the installation location meets environmental conditions for reliable operation.

Source
“max. ambient temperature 40 °C”