Introduces the Unichiller 025w-H chiller and its main features
Specpp. 1-2
Lists performance, temperature, and physical specifications per DIN 12876
p. 2table, 8 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
sound pressure level +/- 4 dB(A)
60 dB(A)
Power supply requirement
220-240V 1~ 50Hz
max. current
15 A
Fuse
16 A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.1/24
Unichiller 025w-H — Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-10...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Cooling power at ambient temperature +20°C at 15°C
2,5 kW
at 0°C
2 kW
at -10°C
1,2 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,4 kg
CO2 equivalent
0,6 t
Circulation pump
Immersion pump
max. delivery
29 l/min
max. delivery pressure
1 bar
Delivery at 0,5 bar
21 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
84 l/h
Consumption at water 15°C, flow 0°C
84 l/h
Consumption at water 15°C, flow -10°C
72 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,8 l
Volume of expansion
1,7 l
Overall dimensions WxDxH **
420x487x579 mm
Net weight
63 kg
** см. сноску в оригинале (в тексте страницы пояснение к ** отсутствует)
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.
Determines the required cooling water supply pressure and the need for a pressure-reducing valve if the site pressure exceeds 6 bar.
Source
“min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar”
Installation
The unit is designed for cooling water intake temperatures between 5°C and 20°C, and below the condensation point the cooling water pipes inside the unit must be insulated.
Defines the acceptable cooling water temperature window and the mandatory insulation measure to prevent condensation inside the unit.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase accordingly.
Allows the engineer to size the cooling water supply and anticipate reduced cooling capacity at higher water temperatures.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.
Guides water quality and material compatibility checks to avoid corrosion or fouling of the cooling water circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must not be used; water mixed with glysantine or ethylene glycol is recommended for freeze protection.
Prevents chemical attack on the pump and specifies an acceptable antifreeze strategy for low-temperature operation.
Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer). It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Conformité
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
Defines the acceptable power supply quality and a prohibited tolerance combination that could damage the unit.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%
Example
-5% voltage and +2% frequency -> not allowed!”
Installation
Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
Ensures the correct mains plug is ordered for the installation site and avoids delays at commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit contains 0.4 kg corresponding to 0.6 t CO2 equivalent.
Informs refrigerant safety classification, charge quantity and CO2-equivalent reporting for site safety and environmental compliance.
Source
“R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,4 kg
CO2 equivalent
0,6 t”
Installation
The unit has protection degree IP20 and must be operated within an ambient temperature range of 5°C to 40°C.
Sets the environmental limits and ingress protection for selecting a suitable installation location.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Selection
The maximum current is 15 A and the fuse rating is 16 A on a 220-240 V single-phase 50 Hz supply.
Used to size the electrical supply circuit and protective device for the chiller.
Source
“Power supply requirement
220-240V 1~ 50Hz
max. current
15 A
Fuse
16 A”
Operation
The minimum filling capacity is 3.8 l and the expansion volume is 1.7 l, which must be respected when filling the bath.
Ensures sufficient heat transfer fluid volume and expansion space to avoid overflow or pump cavitation.
Source
“min. filling capacity
3,8 l
Volume of expansion
1,7 l”
Maintenance
The unit is covered by a 4-year warranty, but registration is required.
Reminds the engineer to register the unit to secure the extended warranty coverage.