Introduces the Unichiller 040T model and its main features
Specpp. 1-2
Performance, temperature, and dimensional data per DIN 12876
p. 2table, 2 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
max. ambient temperature
40 °C
from Serial-No.:
S467100
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-10...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Temperature stability at -10°C
0,5 K
Safety classification
I / NFL
Cooling power at 15°C
4 kW
Cooling power at 0°C
2,5 kW
Cooling power at -10°C
1,1 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
1,9 kg
CO2 equivalent
2,7 t
Circulation pump:
B
max. delivery
26 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
20 l/min
Delivery at 1,0 bar
16 l/min
Delivery at 1,5 bar
12 l/min
Delivery at 2,0 bar
8 l/min
Pump connection
G3/4 male
min. filling capacity
3,5 l
Volume of expansion
3,5 l
Overall dimensions WxDxH **
500x552x1451 mm
Net weight
167 kg
sound pressure level +/- 4 dB(A)
69 dB(A)
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
5,5 A
Fuse (3 phase)
3x10A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires a three-phase 400 V supply with a 3x10 A fuse, so the electrical installation must be sized accordingly.
Determines the mains connection, cable and breaker rating needed at the installation site.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and fit the plug.
Affects site preparation and the bill of materials for commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances must be met simultaneously, and a -5% voltage with +2% frequency combination is explicitly not allowed.
Prevents selecting a supply point whose combined voltage/frequency deviation would violate the permitted operating envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Performance
Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises, so capacity must be derated for warm rooms.
Needed to correctly size the chiller for the actual ambient conditions at the installation site.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit is rated IP20, so it must be installed in a dry, protected location away from water and dust ingress.
Restricts where the chiller can be placed and what environmental protection is required.
Source
“Degree of Protection IP20”
Installation
The minimum ambient temperature is 5 °C and the maximum is 40 °C, defining the allowable room temperature range for installation.
Ensures the installation room stays within the permitted ambient range for reliable operation.
Source
“min. ambient temperature 5 °C”
Safety
The refrigerant is R-449A with GHS classification H280 (pressurised gas), so handling and leak response must follow pressurised-gas precautions.
Informs safety planning, ventilation and refrigerant handling requirements.
Source
“R-449A (A1, H280)”
Conformité
The unit falls under Pressure Equipment Directive Article 4.3, which affects documentation and conformity requirements.
Relevant for regulatory compliance and acceptance documentation at handover.
Source
“Pressure equipment category Art. 4.3 PED”
Conformité
The equipment is EMC class A group 1, so it may cause radio interference in residential areas and is intended for industrial environments.
Determines whether the chiller is suitable for the intended EMC environment and what filtering may be needed.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The pump connection is G3/4 male, so mating hoses and fittings must match this thread.
Ensures correct hydraulic connection hardware is ordered and installed.
Source
“Pump connection G3/4 male”
Operation
The minimum filling capacity is 3.5 l and the expansion volume is 3.5 l, so the external loop must be filled and expansion allowance planned accordingly.
Needed to size the heat transfer fluid charge and expansion volume for the application.
Source
“min. filling capacity 3,5 l”
Maintenance
The unit contains 1.9 kg of R-449A with a CO2 equivalent of 2.7 t, which sets leak-check and reporting obligations.
Determines refrigerant leak inspection intervals and environmental reporting requirements.
Source
“CO2 equivalent 2,7 t”
Operation
The sound pressure level is 69 dB(A) ±4 dB(A), so acoustic planning is needed for the installation location.
Affects placement and any noise mitigation required in the working environment.
Source
“sound pressure level +/- 4 dB(A) 69 dB(A)”
Installation
Space requirements around the unit must be respected, so clearance must be planned per the operating conditions.
Prevents installation in a location without adequate clearance for air flow and service access.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”