Unichiller 045T Datasheet

Sommaire

  1. Overview p. 1
    Introduces the Unichiller 045T chiller unit
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 2 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
max. ambient temperature40 °C
from Serial-No.:S453539

Technical data according to DIN 12876 — Unichiller 045T (Order-No.: 3055.0002.01)

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,5 K
Safety classificationI / NFL
Cooling power at 15°C4,5 kW
Cooling power at 0°C4 kW
Cooling power at -10°C2,7 kW
Cooling power at -20°C1,4 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity2,55 kg
CO2 equivalent3,6 t
Circulation pump:B
max. delivery26 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar20 l/min
Delivery at 1,0 bar16 l/min
Delivery at 1,5 bar12 l/min
Delivery at 2,0 bar8 l/min
Pump connectionG3/4 male
min. filling capacity3,5 l
Volume of expansion3,5 l
Overall dimensions WxDxH **500x552x1451 mm
Net weight183 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)5 A
Fuse (3 phase)3x10A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 40 °C, so the process must stay within these limits.

Defines the usable process temperature window and excludes applications outside it.

Source
“Operating temperature range -20...40 °C”
Performance

Cooling capacity drops from 4.5 kW at 15 °C to 1.4 kW at -20 °C, so sizing must be based on the actual setpoint.

Prevents undersizing when the process runs at low temperatures.

Source
“at 15°C 4,5 kW at 0°C 4 kW at -10°C 2,7 kW at -20°C 1,4 kW”
Installation

The unit requires a 400 V three-phase supply with a 3x10 A fuse, so the electrical installation must be prepared accordingly.

Determines the site power feed, protection device and cable sizing.

Source
“Power supply (3 Phase) 400V 3~ 50Hz max. current (3 Phase) 5 A Fuse (3 phase) 3x10A”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the plug and connection.

Affects site electrical preparation and commissioning scope.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.

Restricts the installation environment and rules out washdown or outdoor locations.

Source
“Degree of Protection IP20”
Installation

Ambient temperature must stay between 5 and 40 °C, so the room ventilation and location must be checked before installation.

Ensures the refrigeration unit and electronics operate within their rated ambient envelope.

Source
“min. ambient temperature 5 °C”
Performance

Cooling capacity is stated at 20 °C room temperature and may fall if ambient rises, so hot rooms require derating.

Avoids overestimating available cooling in warm plant rooms.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The refrigerant is R-449A with GWP 1397 and a CO2 equivalent of 3.6 t, so refrigerant reporting and leak checks apply.

Triggers F-gas/CO2-equivalent documentation and service obligations.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280) Global Warming Potential (GWP) 1397 Refrigerant quantity 2,55 kg CO2 equivalent 3,6 t”
Safety

The refrigerant is classified A1 with H280, so it is non-flammable but stored under pressure and requires pressure-system awareness.

Informs risk assessment and handling procedures for the refrigeration circuit.

Source
“R-449A (A1, H280)”
Conformité

The pressure equipment falls under Article 4.3 PED, so it is classified according to the Pressure Equipment Directive.

Determines the applicable pressure-equipment compliance route.

Source
“Pressure equipment category Art. 4.3 PED”
Conformité

The unit is EMC Class A Group 1 to EN55011, so it is intended for industrial environments and may cause interference in residential areas.

Affects installation location and EMC coordination with sensitive equipment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and opposite-sign deviations are not allowed.

Defines the acceptable supply quality for reliable operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation

The pump delivers 26 l/min maximum and 8 l/min at 2.0 bar, so the external circuit must be checked against the pump curve.

Ensures the external loop pressure drop is compatible with the circulation pump.

Source
“max. delivery 26 l/min max. delivery pressure 3 bar”
Maintenance

Minimum filling capacity is 3.5 l and expansion volume is 3.5 l, so the heat-transfer fluid charge and expansion allowance must be planned.

Prevents underfilling and overflow/expansion issues during operation.

Source
“min. filling capacity 3,5 l Volume of expansion 3,5 l”