Introduces the Unichiller 050w chiller model and its main features
Specpp. 1-2
Performance, temperature, and dimensional data per DIN 12876
p. 2table, 13 rows
p. 1table, 36 rows
Technical data according to DIN 12876
Parameter
Value
min. filling capacity
18 l
Volume of expansion
48 l
Overall dimensions WxDxH **
740x1160x1050 mm
Net weight
285 kg
sound pressure level +/- 4 dB(A)
62 dB(A)
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
6,5 A
Fuse (3 phase)
3x10A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S520192
** Please respect space requirements. See operating conditions at www.huber-online.com
Unichiller 050w — Technical data according to DIN 12876
Parameter
Value
Order-No.
3038.0056.01
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at ambient temperature +20°C
at 20°C
5 kW
at 0°C
4,2 kW
at -10°C
3 kW
at -20°C
1,8 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,85 kg
CO2 equivalent
1,2 t
Circulation pump
E1
max. delivery
48 l/min
max. delivery pressure
3,4 bar
Delivery at 0,3 bar
46 l/min
Delivery at 0,5 bar
44 l/min
Delivery at 1,0 bar
40 l/min
Delivery at 1,5 bar
33 l/min
Delivery at 2,0 bar
27 l/min
Delivery at 2,5 bar
20 l/min
Delivery at 3,0 bar
11 l/min
Pump connection
G1 1/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 20°C
240 l/h
Consumption at water 15°C, flow 0°C
216 l/h
Consumption at water 15°C, flow -10°C
167 l/h
Consumption at water 15°C, flow -20°C
137 l/h
min. cooling water differential pressure
0,4 bar
max. cooling water pressure
6 bar
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires cooling water supply with a differential pressure of at least 0.4 bar and no more than 6 bar.
To ensure adequate cooling water flow and avoid damage to the refrigeration circuit.
Source
“min. cooling water differential pressure 0,4 bar”
Installation
The unit must be connected to a three-phase 400 V 50 Hz power supply with a 3x10 A fuse.
To verify site power availability and correct breaker sizing before installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices with current consumption less than 63 A are delivered with a cable but without a plug, so the customer must provide a suitable plug.
To plan for electrical termination and avoid delays during commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The cooling water inlet temperature must be between 5 °C and 20 °C; if it falls below the condensation point, the cooling water pipes inside the unit must be insulated.
To prevent condensation and corrosion inside the unit when using cold cooling water.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.
To correctly size the cooling water supply and avoid unexpected performance loss.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
To select suitable water quality and avoid corrosion or material degradation.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité
Voltage tolerance is ±5% with a simultaneous frequency tolerance of ±2%, but the combination of -5% voltage and +2% frequency is not allowed.
To assess site power quality and ensure the unit operates within its specified tolerances.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant R-449A has a Global Warming Potential of 1397 and a CO2 equivalent of 1.2 t, which must be considered for environmental compliance.
To evaluate refrigerant regulations and potential leak reporting obligations.
Source
“Global Warming Potential (GWP) 1397”
Installation
The unit has a protection degree of IP20, so it must be installed in a dry indoor environment.
To ensure the installation location meets environmental protection requirements.
Source
“Degree of Protection IP20”
Operation
The ambient temperature for operation must be between 5 °C and 40 °C.
To verify that the installation site's ambient conditions are within the allowed range.
Source
“min. ambient temperature 5 °C”
Installation
The unit requires minimum filling capacity of 18 l and has an expansion volume of 48 l.
To prepare the correct amount of heat transfer fluid and expansion tank capacity.
Source
“min. filling capacity 18 l”
Performance
The circulation pump delivers a maximum flow of 48 l/min and a maximum pressure of 3.4 bar.
To check whether the pump performance matches the external circuit's pressure drop and flow requirements.
Source
“max. delivery 48 l/min”
Installation
The pump connection is G1 1/4 male and the cooling water connection is G1/2 male.
To procure correct fittings and hoses for installation.
Source
“Pump connection G1 1/4 male”
Conformité
The equipment is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011.
To assess electromagnetic compatibility in the installation environment and plan for potential interference.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”