Sommaire

  1. Overview p. 1
    Introduces the Unichiller 055T-H chiller and its main features
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 13 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
min. filling capacity5 l
Volume of expansion7,5 l
Overall dimensions WxDxH **640x937x1606 mm
Net weight283 kg
sound pressure level +/- 4 dB(A)74 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)12 A
Fuse (3 phase)3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.1/26

** Please respect space requirements. See operating conditions at www.huber-online.com

Unichiller 055T-H — Technical data according to DIN 12876

ParameterValue
Order-No.3015.0095.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,2 K
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C5,5 kW
Cooling power at 0°C2,3 kW
Cooling power at -10°C0,8 kW
Cooling power at -20°C0,2 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,95 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Circulation pump:C3
max. delivery100 l/min
max. delivery pressure5,6 bar
Delivery at 0,3 bar97 l/min
Delivery at 1,0 bar90 l/min
Delivery at 1,5 bar84 l/min
Delivery at 2,0 bar77 l/min
Delivery at 2,5 bar71 l/min
Delivery at 3,0 bar64 l/min
Delivery at 3,5 bar57 l/min
Delivery at 4,0 bar48 l/min
Delivery at 4,5 bar39 l/min
Delivery at 5,0 bar27 l/min
Delivery at 5,5 bar8 l/min
Pump connectionG1 1/4 male

Engineering Facts

13 facts · every claim linked to a page
Selection

The chiller is intended only for externally closed applications, so the user circuit must be closed.

Determines whether the unit can be used with an open bath or only with a closed external loop.

Source
“For externally closed applications.”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects installation and safety planning.

Flammable refrigerant classification requires appropriate site safety and handling measures.

Source
“R-290 (A3, H220)”
Installation

The unit is air-cooled, so sufficient air flow and ambient temperature limits must be respected.

Air-cooled condensers need adequate ventilation and ambient conditions for rated performance.

Source
“air-cooled, natural”
Performance

Cooling capacity drops strongly at lower temperatures, from 5.5 kW at 15 °C to 0.2 kW at -20 °C.

The process load must be checked against the actual cooling capacity at the required operating temperature.

Source
“at 15°C 5,5 kW at 0°C 2,3 kW at -10°C 0,8 kW at -20°C 0,2 kW”
Installation

The unit requires a three-phase 400 V 3~ 50 Hz supply with a 3x16 A fuse.

Electrical supply and protection must be provisioned before installation.

Source
“400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.

The installer must plan and provide the correct plug and connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The allowable ambient temperature range is 5 °C to 40 °C, and cooling capacity may fall above 20 °C.

Site ambient conditions must be within limits and performance derating considered.

Source
“max. ambient temperature 40 °C”
Performance

Output data are valid at 20 °C room temperature, and higher ambient temperature may reduce cooling capacity.

Selection calculations should account for actual ambient temperature at the installation site.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but not all combinations are allowed.

The power quality at the site must be checked to avoid invalid supply combinations.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité

The equipment is EMC classified as Class A, Group 1, which has implications for use in residential environments.

Class A equipment may require additional EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The unit has IP20 protection, so it must be protected from water and dust ingress.

The installation location must avoid moisture and conductive dust.

Source
“Degree of Protection IP20”
Maintenance

The minimum filling capacity is 5 l and expansion volume is 7.5 l, which must be considered when filling the loop.

Insufficient fill volume or expansion space can cause operational problems.

Source
“min. filling capacity 5 l”
Installation

The unit is heavy at 283 kg and large at 640x937x1606 mm, requiring adequate floor space and handling equipment.

Transport, floor loading and clearance must be planned before installation.

Source
“640x937x1606 mm”