Unichiller 070T-H Technical Data

Sommaire

  1. Overview p. 1
    Product name and general introduction to the chiller unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range-10...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,2 K
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C7 kW
Cooling power at 0°C4 kW
Cooling power at -10°C2,3 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pump:C3
max. delivery84 l/min
max. delivery pressure5,6 bar
Delivery at 0,33 bar81 l/min
Delivery at 0,5 bar79 l/min
Delivery at 1,0 bar74 l/min
Delivery at 1,5 bar69 l/min
Delivery at 2,0 bar64 l/min
Delivery at 2,5 bar58 l/min
Delivery at 3,0 bar52 l/min
Delivery at 3,5 bar46 l/min
Delivery at 4,0 bar39 l/min
Delivery at 4,5 bar30 l/min
Delivery at 5,0 bar21 l/min
Delivery at 5,5 bar7 l/min
Pump connectionG1 1/4 male
min. filling capacity5 l
Volume of expansion7,5 l
Overall dimensions WxDxH **600x790x1614 mm

** см. сноску в оригинале (в тексте страницы пояснение к ** отсутствует)

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -10 to 100 °C, so it must be matched to the process temperature window.

Defines whether the unit can cover the required process temperatures before selection.

Source
“Operating temperature range -10...100 °C”
Performance

Cooling capacity drops from 7 kW at 15 °C to 2.3 kW at -10 °C, so the duty point must be checked at the actual operating temperature.

Prevents oversizing or undersizing when the process runs at low temperature.

Source
“at 15°C 7 kW at 0°C 4 kW at -10°C 2,3 kW”
Installation

The unit requires a 400 V three-phase 50 Hz supply, so site power must be verified before installation.

Determines electrical infrastructure and whether a suitable three-phase feed is available.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without plug, so the customer must provide the final connection.

Affects electrical installation planning and who supplies the plug or terminal connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively but must not be combined in opposite directions, so supply quality must be checked.

Avoids invalid supply conditions that could damage the compressor or void compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation

The unit is rated for ambient temperatures from 5 to 40 °C, so the installation room must stay within this range.

Ambient conditions directly affect cooling performance and reliable operation.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity is valid at 20 °C room temperature and may drop at higher ambient, so performance must be derated for warm rooms.

Prevents selecting a unit that cannot meet duty in a warm plant room.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit is air-cooled, so adequate air flow and heat rejection space must be provided at the installation site.

Air-cooled units need ventilation and clearance, which affects room layout.

Source
“Refrigeration machine air-cooled, CFC- and HCFC-free”
Safety

The refrigerant is R-449A with GHS classification H280, so refrigerant handling and ventilation precautions apply.

Requires awareness of pressurized gas hazards during installation and service.

Source
“R-449A (A1, H280)”
Conformité

The unit is classified Class A, Group 1 for electromagnetic disturbance, so it is intended for industrial environments.

Class A equipment may not be suitable for residential or light-commercial EMC environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The degree of protection is IP20, so the unit must be installed in a dry, protected location.

Prevents installation in wet or dusty areas where ingress protection is required.

Source
“Degree of Protection IP20”
Installation

The pump connection is G1 1/4 male, so mating hoses and fittings must match this thread.

Ensures correct hydraulic connection and avoids adapters or leaks.

Source
“Pump connection G1 1/4 male”
Operation

The minimum filling capacity is 5 l and expansion volume is 7.5 l, so the external loop volume must be considered.

Affects heat transfer fluid charge and expansion tank sizing for the application.

Source
“min. filling capacity 5 l Volume of expansion 7,5 l”
Maintenance

The sound pressure level is 66 dB(A) ±4 dB(A), so acoustic planning may be needed in occupied areas.

Noise limits may dictate location or acoustic treatment.

Source
“sound pressure level +/- 4 dB(A) 66 dB(A)”