Introduces the Unichiller 100-H chiller and its main features
Specpp. 1-2
Performance, temperature, cooling capacity and dimensional data per DIN 12876
p. 2table, 9 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
sound pressure level +/- 4 dB(A)
58 dB(A)
Power supply (3 Phase)
460V 3~ 60Hz
max. current (3 Phase)
13,5 A
Fuse (3 phase)
3x16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S438249
Technical data according to DIN 12876 — Unichiller 100-H (Order-No.: 3059.0006.01)
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at ambient temperature +20°C
at 100°C
10 kW
at 20°C
10 kW
at 0°C
9,7 kW
at -10°C
6,7 kW
at -20°C
4,6 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
1,1 kg
CO2 equivalent
1,5 t
Circulation pump:
E1
max. delivery
56 l/min
max. delivery pressure
3,8 bar
Delivery at 0,3 bar
53 l/min
Delivery at 0,5 bar
51 l/min
Delivery at 1,0 bar
46 l/min
Delivery at 1,5 bar
40 l/min
Delivery at 2,0 bar
34 l/min
Delivery at 2,5 bar
27 l/min
Delivery at 3,0 bar
18 l/min
Pump connection
G1 1/4 male
min. filling capacity
18 l
Volume of expansion
48 l
Overall dimensions WxDxH **
740x1160x1165 mm
Net weight
284 kg
** см. сноску в оригинале (на странице не раскрыта).
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires a three-phase 460 V 60 Hz supply with a 3x16 A fuse, so the site electrical infrastructure must be provisioned accordingly.
Determines the electrical supply, breaker/fuse rating and cabling needed at the installation site.
Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation
An external branch circuit protection according to UL 489 is required, so a suitable UL 489 listed breaker must be installed upstream.
Affects the electrical panel design and compliance of the installation.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and wire the plug or connection.
Impacts site wiring preparation and the need to source a suitable plug/terminal connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The unit must be installed in an environment between 5 °C and 40 °C ambient, so the room climate must be kept within this range.
Room conditions outside this range can prevent proper operation and affect cooling performance.
Source
“min. ambient temperature 5 °C”
Performance
Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises, so capacity margins must account for the actual room temperature.
Prevents undersizing the chiller when the installation room runs warmer than 20 °C.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.
Restricts where the chiller can be placed and what environmental protection is needed.
Source
“Degree of Protection IP20”
Installation
The pump connection is G1 1/4 male, so hoses and fittings must match this thread size.
Determines the hose couplings and adapters required for the external circuit.
Source
“Pump connection G1 1/4 male”
Selection
The unit is intended for externally closed applications, so the process loop must be a closed circuit.
Confirms the chiller is not suitable for open-loop or once-through cooling setups.
Source
“For externally closed applications.”
Selection
The minimum filling capacity is 18 l and expansion volume is 48 l, so the external circuit volume must be considered when sizing the system.
Helps verify the external loop volume and heat transfer fluid quantity needed.
Source
“min. filling capacity 18 l”
Safety
The refrigerant is R-449A with GHS classification H280, so refrigerant handling and ventilation precautions apply.
Affects safety planning, room ventilation and refrigerant handling procedures.
Source
“R-449A (A1, H280)”
Conformité
The equipment is classified Class A, Group 1 for electromagnetic disturbance, so it may require EMC measures in residential or sensitive environments.
Class A equipment is not intended for residential use without additional EMC precautions.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but not simultaneously in the unfavorable combination, so supply quality must be checked.
Ensures the site power quality stays within the allowed envelope to avoid damage or malfunction.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation
The chiller has a sound pressure level of 58 dB(A) ±4 dB(A), so acoustic planning is needed for the installation room.
Relevant for noise-sensitive environments and room placement.
Source
“sound pressure level +/- 4 dB(A) 58 dB(A)”
Maintenance
The refrigeration machine is air-cooled, so adequate airflow and clearance around the unit are required for heat rejection.
Air-cooled units need unobstructed ventilation and space for maintenance access.