Induztr
Résumé technique Visionneuse de documents Modèles associés

Unichiller 110Tw Datasheet

Sommaire

  1. Overview p. 1
    Introduces the Unichiller 110Tw chiller model and its main features
  2. Spec p. 2
    Technical data per DIN 12876 including performance and dimensions
  3. p. 2 table, 16 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow -10°C89 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity6,5 l
Volume of expansion5,5 l
Overall dimensions WxDxH **600x600x1450 mm
Net weight225 kg
sound pressure level +/- 4 dB(A)65 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)13,5 A
Fuse (3 phase)3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S589077

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller 110Tw (Order-No.: 3027.0067.01)

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at 15°C11 kW
Cooling power at 0°C6 kW
Cooling power at -10°C2,7 kW
Cooling power at -20°C2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,75 kg
CO2 equivalent2,4 t
UN-numberUN 2857
Circulation pumpC3
max. delivery96 l/min
max. delivery pressure5,6 bar
Delivery at 0,5 bar92 l/min
Delivery at 1,0 bar86 l/min
Delivery at 1,5 bar81 l/min
Delivery at 2,0 bar75 l/min
Delivery at 2,5 bar68 l/min
Delivery at 3,0 bar62 l/min
Delivery at 3,5 bar54 l/min
Delivery at 4,0 bar46 l/min
Delivery at 4,5 bar37 l/min
Delivery at 5,0 bar24 l/min
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 15°C400 l/h
Consumption at water 15°C, flow 0°C191 l/h

Значения приведены дословно по тексту страницы; таблица не является матрицей конфигураций, поэтому транспонирование не применялось.

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires a cooling water supply with a differential pressure between 1 and 6 bar.

Determines whether the site's cooling water supply is adequate and whether a pressure reducer is needed.

Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation

The unit is designed for cooling water intake between 5°C and 20°C, and below the condensation point the cooling water pipes inside the unit must be insulated.

Prevents condensation and corrosion/electrical issues when the site supplies cold cooling water.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.

The published 11 kW at 15°C is only valid under the stated conditions; warmer site water reduces available capacity.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility to avoid corrosion and fouling of the heat exchanger.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Installation

The three-phase unit draws up to 13.5 A and requires a 3x16 A fuse on a 400 V 3~ 50 Hz supply.

Sizing of the electrical supply, cable and protection device for installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz max. current (3 Phase) 13,5 A Fuse (3 phase) 3x16A”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and the bill of materials for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Conformité

Voltage and frequency tolerances are only valid in combination: ±5% voltage with ±2% frequency, and some combinations are not allowed.

Ensures the site power quality is within the permitted envelope before installation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The refrigerant is R-449A with GHS classification A1/H280 and a GWP of 1397, with 1.75 kg charge equal to 2.4 t CO2 equivalent.

Relevant for refrigerant handling, leak detection, F-gas reporting and room ventilation requirements.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280) Global Warming Potential (GWP) 1397 Refrigerant quantity 1,75 kg CO2 equivalent 2,4 t”
Safety

The unit is classified as pressure equipment category I and has protection degree IP20.

Informs installation environment requirements and electrical protection level.

Source
“Pressure equipment category I Degree of Protection IP20”
Operation

The unit must be operated within an ambient temperature range of 5°C to 40°C.

Defines the allowable room conditions for reliable operation and performance.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The unit needs a minimum filling capacity of 6.5 l and has an expansion volume of 5.5 l.

Determines the minimum heat transfer fluid volume and expansion tank sizing for the external circuit.

Source
“min. filling capacity 6,5 l Volume of expansion 5,5 l”
Installation

The pump connections are G1 1/4 male and the cooling water connections are G3/4 male.

Needed to select matching hoses, adapters and fittings for installation.

Source
“Pump connection G1 1/4 male Cooling water connection G3/4 male”
Performance

The circulation pump delivers a maximum of 96 l/min at 5.6 bar, with flow decreasing as pressure rises.

Used to check whether the external circuit's pressure drop is compatible with the available pump curve.

Source
“max. delivery 96 l/min max. delivery pressure 5,6 bar”
Operation

The unit is intended for externally closed applications.

Prevents misapplication with open baths or circuits that would not suit the expansion tank design.

Source
“For externally closed applications.”