Selection
The chiller's cooling capacity drops as ambient temperature rises, so size the unit for the actual room conditions.
Cooling capacity ratings are referenced to 20 °C ambient; higher ambient reduces available capacity and may require a larger unit.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
Three-phase units under 63 A are supplied with cable but no plug, so the installer must provide a suitable plug or hardwiring.
Affects electrical installation planning and bill of materials for the 400 V three-phase supply.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances must be met simultaneously, so verify the site supply before commissioning.
Prevents invalid supply combinations that could damage the compressor or pump.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
The unit is IP20, so it must be installed in a dry indoor location protected from water ingress.
IP20 is not water-protected; placement in wet or washdown areas is not permitted.
Source
“Degree of Protection
IP20”
Installation
The unit is air-cooled and requires adequate space for airflow, so respect the stated space requirements.
Insufficient clearance reduces cooling performance and can cause overheating.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Operation
The unit is intended only for externally closed applications, so do not use it with an open bath.
Using it in an open configuration risks overflow and incorrect operation.
Source
“For externally closed applications.”
Operation
The expansion tank is atmospheric open, so the installation must account for fluid level and venting.
An open expansion tank requires correct fluid filling and level monitoring.
Source
“Housing, atmospheric open expansion tank
and copper soldered evaporator made of stainless steel.”
Safety
The refrigerant is R-449A with GHS H280, so handling and disposal must follow pressurized-gas safety rules.
H280 indicates a pressurized gas hazard; service personnel must be qualified and follow local regulations.
Source
“R-449A (A1, H280)”
Safety
The unit has adjustable overtemperature protection per DIN 12876, so set it to protect the application.
Proper setting of the overtemperature cut-out protects the process and the equipment.
Source
“With
adjustable overtemperature protection according to DIN 12876.”
Performance
The circulation pump delivers 210 l/min at zero head but only 31 l/min at 4.5 bar, so match the external circuit pressure drop.
External loop pressure drop determines actual flow; undersized piping will reduce cooling performance.
Source
“max. delivery
210 l/min”
Performance
Cooling capacity falls from 20 kW at 15 °C to 3 kW at -20 °C, so verify capacity at the actual operating temperature.
Selecting on the 15 °C rating alone will undersize the unit for low-temperature duty.
Source
“at 15°C
20 kW”
Conformité
The unit is EMC Class A Group 1, so it may cause interference in residential environments.
Class A equipment is intended for industrial use and may require additional filtering in sensitive areas.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
The refrigerant has a GWP of 1397, so leak checks and record-keeping are required under F-gas rules.
High GWP refrigerant triggers periodic leak inspection and documentation obligations.
Source
“Global Warming Potential (GWP)
1397”