Installation
The chiller requires a three-phase 400 V supply and is delivered without a plug, so the site must provide a suitable 3-phase connection.
Determines electrical installation scope and whether a plug must be sourced separately.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Cooling water inlet must be kept between 5 °C and 20 °C, and pipes must be insulated if water can drop below the dew point.
Defines the required facility cooling water supply conditions and condensation protection.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
A minimum cooling water differential pressure of 1 bar is required for rated performance.
Ensures the facility cooling water loop can deliver adequate flow through the unit.
Source
“min. cooling water differential pressure”
Installation
Cooling water pressure must not exceed 6 bar.
Prevents overpressure damage to the cooling water circuit.
Source
“max. cooling water pressure”
Installation
The unit needs at least 15 l filling capacity and has a 22 l expansion volume, so the external loop volume must be considered.
Affects heat transfer fluid volume planning and expansion tank sizing.
Source
“min. filling capacity”
Installation
Ambient temperature must stay between 5 °C and 40 °C for safe operation.
Constrains the installation location and room ventilation requirements.
Source
“min. ambient temperature”
Installation
The unit is IP20, so it must be installed in a dry indoor location protected from water ingress.
Limits where the chiller can be placed and what environmental protection is needed.
Source
“Degree of Protection”
Installation
The unit is water-cooled and uses R-449A refrigerant with GWP 1397, which affects environmental compliance and leak management.
Relevant for refrigerant handling, leak detection and environmental reporting.
Source
“Refrigerant (ASHRAE, GHS)”
Operation
Cooling capacity drops as cooling water temperature rises, and cooling water flow may need to increase.
Prevents undersizing the cooling water supply and unexpected performance loss.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Operation
Voltage and frequency tolerances are ±5% and ±2% respectively, but not all combinations are allowed.
Ensures the power supply quality is within the unit's acceptable operating envelope.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Operation
The pump delivers 210 l/min maximum and 4.7 bar maximum pressure, with flow falling at higher pressure.
Used to check whether the external circuit resistance is compatible with the pump curve.
Source
“max. delivery”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.
Confirms built-in safety protection for unattended or critical temperature control.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Conformité
The equipment is EMC Class A Group 1, which may require additional filtering in residential environments.
Affects EMC planning and suitability for different installation environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and material compatibility to avoid corrosion and fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”