Introduces the Unichiller 350T chiller and its main features
Specpp. 1-2
Lists performance, temperature, and dimensional data per DIN 12876
p. 2table, 6 rows
p. 1table, 34 rows
Technical data according to DIN 12876
Parameter
Value
Fuse (3 phase)
3x32A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S602655
Technical data according to DIN 12876 — Unichiller 350T (Order-No.: 3021.0007.01)
Parameter
Value
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Temperature stability at -10°C
0,5 K
Safety classification
I / NFL
Cooling power at 15°C
35 kW
Cooling power at 0°C
23 kW
Cooling power at -10°C
14 kW
Cooling power at -20°C
8 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
8 kg
CO2 equivalent
11,2 t
Circulation pump:
D3
max. delivery
210 l/min
max. delivery pressure
4,7 bar
Delivery at 1,1 bar
202 l/min
Delivery at 1,5 bar
193 l/min
Delivery at 2,0 bar
179 l/min
Delivery at 2,5 bar
159 l/min
Delivery at 3,0 bar
140 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
77 l/min
Delivery at 4,5 bar
31 l/min
Pump connection
G1 1/4 male
min. filling capacity
14 l
Volume of expansion
39 l
Overall dimensions WxDxH **
905x2172x1900 mm
Net weight
640 kg
Power supply (3 Phase)
460V 3~ 60Hz
max. current (3 Phase)
27 A
** см. сноску в оригинале (Overall dimensions WxDxH **)
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires a 460 V three-phase 60 Hz supply with a 3x32 A fuse, so the site electrical infrastructure must be sized accordingly.
Determines the electrical supply, cable sizing and branch protection needed at the installation site.
Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation
External branch circuit protection according to UL 489 is mandatory, so a suitable UL-listed breaker must be provided by the installer.
Affects the electrical panel design and compliance of the installation.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so the customer must provide the final connection.
Impacts site wiring and the need to procure a suitable plug or hard-wired connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The unit must be installed in an ambient temperature between 5 °C and 40 °C, so the room climate must be controlled.
Room conditions outside this range can prevent proper operation and may void performance claims.
Source
“min. ambient temperature 5 °C”
Performance
Cooling capacity is rated at 20 °C room temperature and drops if ambient temperature rises, so capacity margins must account for the actual room temperature.
Prevents undersizing the chiller when the installation room is warmer than 20 °C.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
Voltage and frequency tolerances are ±5 % and ±2 % but must not be combined in the unfavorable direction, so the supply quality must be verified.
Ensures the chiller is not connected to a supply that could damage the compressor or electronics.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions are required.
Affects safety planning, room ventilation and service procedures for the refrigeration circuit.
Source
“R-449A (A1, H280)”
Conformité
The CO2 equivalent of the refrigerant charge is 11.2 t, which triggers F-gas record-keeping and leak-check obligations in many jurisdictions.
Determines regulatory obligations for refrigerant monitoring and periodic leak checks.
Source
“CO2 equivalent 11,2 t”
Installation
The unit is IP20, so it must be installed indoors protected from water and dust ingress.
Restricts installation locations and requires a clean, dry environment.
Source
“Degree of Protection IP20”
Installation
The chiller is air-cooled, so adequate air flow and clearance around the unit are needed for heat rejection.
Poor ventilation will reduce cooling capacity and can cause overheating.
Source
“Refrigeration machine air-cooled, CFC- and HCFC-free”
Installation
The overall dimensions are 905 x 2172 x 1900 mm and the net weight is 640 kg, so floor loading and access routes must be checked.
Affects rigging, doorway clearance and structural floor capacity at the installation site.
Source
“905x2172x1900 mm”
Installation
The pump connection is G1 1/4 male, so mating hoses and fittings must match this thread.
Ensures correct hydraulic connection and avoids adapters or leaks.
Source
“Pump connection G1 1/4 male”
Operation
The minimum filling capacity is 14 l and the expansion volume is 39 l, so the external loop volume must be compatible with the expansion tank.
Prevents overflow or insufficient expansion volume in externally closed applications.
Source
“min. filling capacity 14 l”
Conformité
The equipment is classified Class A, Group 1 for electromagnetic disturbance, so it may require EMC measures in residential or sensitive environments.
Affects EMC planning and suitability for certain installation environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”