Unichiller 500Tw-H Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 8 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Overall dimensions WxDxH **1000x1100x1635 mm
Net weight634 kg
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/20

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ХарактеристикаUnichiller 500Tw-H
Order-No.3030.0014.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C50 kW
Cooling power at 0°C30 kW
Cooling power at -10°C24 kW
Cooling power at -20°C14 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pumpD3
max. delivery234 l/min
max. delivery pressure4,9 bar
Delivery at 1,3 bar197 l/min
Delivery at 1,5 bar192 l/min
Delivery at 2,0 bar178 l/min
Delivery at 2,5 bar158 l/min
Delivery at 3,0 bar137 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar86 l/min
Delivery at 4,5 bar40 l/min
Pump connectionG1 1/4 male
Cooling water connectionG1 1/4 male
Consumption at water 15°C, flow 15°C2040 l/h
min. cooling water differential pressure1,5 bar
max. cooling water pressure6 bar
min. filling capacity12,7 l
Volume of expansion7 l

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires a three-phase 400 V supply and is delivered without a plug, so the site must provide a suitable 63 A-or-less three-phase connection and local isolation.

Determines electrical supply, cable termination and switchgear needed at installation.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Cooling water must be supplied at 5–20 °C inlet with at least 1.5 bar differential pressure and no more than 6 bar, so the facility water system must be checked before commissioning.

Cooling water conditions directly govern achievable cooling capacity and safe operation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water is colder than the ambient dew point, the cooling water pipes inside the unit must be insulated to prevent condensation.

Prevents condensation, corrosion and electrical risk in low-temperature cooling water installations.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Rated cooling capacity is only valid at 20 °C room temperature and 15 °C cooling water with 1.5 bar differential pressure, so higher water temperatures will reduce capacity and may require more flow.

Prevents overestimating cooling capacity when site conditions differ from the rating basis.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1,5 bar differential pressure between cooling water inlet and -outlet.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, brass, PA, PPE, PTFE and EPDM, so the facility water must be compatible with these materials.

Guides water quality and material compatibility checks to avoid corrosion or fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité

Voltage and frequency tolerances are ±5 % and ±2 % respectively, but they may not be combined in opposite directions, so the site supply must be verified against this rule.

Defines acceptable power quality limits for warranty and reliable operation.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The unit is classified IP20, so it must be installed in a dry indoor location protected from water ingress and dust.

Restricts installation environment and enclosure protection requirements.

Source
“Degree of Protection IP20”
Installation

Ambient temperature must stay between 5 °C and 40 °C, so the installation room needs adequate ventilation and heating.

Out-of-range ambient temperature affects performance, reliability and safety.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection

The refrigeration circuit uses R-449A with GWP 1397, which matters for refrigerant leakage reporting and environmental compliance.

Refrigerant type and GWP affect regulatory obligations and service planning.

Source
“R-449A (A1, H280)”
Selection

The pump delivers 234 l/min maximum at 4.9 bar, but flow falls to 40 l/min at 4.5 bar, so system pressure drop must be matched to the pump curve.

Ensures the external circuit resistance does not starve flow and reduce heat transfer.

Source
“max. delivery 234 l/min max. delivery pressure 4,9 bar”
Installation

Pump and cooling water connections are G1 1/4 male, so mating hoses and couplings must be ordered accordingly.

Determines connection hardware needed for piping and commissioning.

Source
“Pump connection G1 1/4 male Cooling water connection G1 1/4 male”
Operation

The unit is intended for externally closed applications, so the external loop must be closed and not open to atmosphere.

Prevents incorrect application that could cause evaporation, contamination or poor temperature control.

Source
“For externally closed applications.”
Maintenance

The minimum filling capacity is 12.7 l with 7 l expansion volume, so the external loop and expansion tank must be sized to hold this volume.

Ensures sufficient heat transfer fluid volume and expansion space for safe operation.

Source
“min. filling capacity 12,7 l Volume of expansion 7 l”
Safety

The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application to protect the process and equipment.

Provides a safety limit that must be configured correctly during commissioning.

Source
“With adjustable overtemperature protection according to DIN 12876.”