Product name and general introduction to the Unichiller 800Tw
Specp. 2
Technical data according to DIN 12876 standard
p. 1table, 34 rows
Technical data according to DIN 12876
Characteristic
Value
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at 40°C
80 kW
Cooling power at 15°C
80 kW
Cooling power at 0°C
60 kW
Cooling power at -10°C
40 kW
Cooling power at -20°C
20 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Circulation pump
D3
max. delivery
196 l/min
max. delivery pressure
5 bar
Delivery at 1,0 bar
175 l/min
Delivery at 2,0 bar
150 l/min
Delivery at 3,0 bar
122 l/min
Delivery at 4,0 bar
74 l/min
Pump connection
G1 1/4 male
Cooling water connection
G1 1/4 male
Consumption at water 15°C, flow 15°C
3480 l/h
min. cooling water differential pressure
2,5 bar
max. cooling water pressure
6 bar
min. filling capacity
12,7 l
Volume of expansion
7 l
Overall dimensions WxDxH **
1000x1600x1620 mm
Power supply (3 Phase)
400V 3~ 50Hz
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
** Overall dimensions WxDxH
Engineering Facts
13 facts · every claim linked to a page
Installation
The chiller requires cooling water supply pressure between 2.5 bar minimum differential and 6 bar maximum, so the site water system must be verified before installation.
Insufficient differential pressure will prevent rated cooling performance and may trip the unit.
Source
“min. cooling water differential pressure 2,5 bar”
Installation
Cooling water inlet temperature must be kept between 5°C and 20°C, and pipes must be insulated if water can drop below the condensation point.
Out-of-range cooling water temperature invalidates the published performance data and risks condensation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Published cooling capacity is only valid at 20°C room temperature, 15°C cooling water inlet and 2.5 bar differential pressure.
Selection must account for actual site conditions, since higher cooling water temperature reduces capacity.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 2,5 bar differential pressure between cooling water inlet and -outlet.”
Installation
The unit is water-cooled and requires 3480 l/h cooling water flow at 15°C, which drives the sizing of the facility water supply.
Cooling water flow capacity must be confirmed before committing to this chiller.
Source
“Consumption at water 15°C, flow 15°C 3480 l/h”
Installation
The chiller is a three-phase 400 V device with current likely above 63 A, so it is supplied without cable and without plug.
Electrical installation must include customer-supplied cabling and a suitable disconnect/protection device.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Conformité
Voltage and frequency tolerances are coupled: ±5% voltage is only allowed with ±2% frequency, and opposite-sign deviations are not permitted.
Site power quality must be checked against this combined tolerance to avoid invalidating operation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation requirements apply.
Refrigerant safety classification affects room ventilation and service procedures.
Source
“R-449A (A1, H280)”
Selection
The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.
IP20 rules out washdown or outdoor installation and constrains the installation location.
Source
“Degree of Protection IP20”
Installation
Ambient temperature must stay between 5°C and 40°C for correct operation.
The installation room must be conditioned to keep ambient temperature within this range.
Source
“min. ambient temperature 5 °C”
Installation
The unit requires a minimum filling capacity of 12.7 l and has 7 l expansion volume, which matters for heat transfer fluid selection and filling.
Fluid volume and expansion must be planned to avoid overflow or low-level alarms.
Source
“min. filling capacity 12,7 l”
Installation
The chiller is intended for externally closed applications, which constrains the hydraulic circuit design.
Open-loop or unpressurized external circuits are not supported by this configuration.
Source
“For externally closed applications.”
Maintenance
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Water chemistry must be compatible with these materials to avoid corrosion and fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
The unit requires significant floor space of 1000x1600x1620 mm, and space requirements must be respected.
Access and clearance for service must be planned around the stated dimensions.
Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”