Unichiller GL 150w Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and general description of the chiller
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 1 table, 27 rows

Unichiller GL 150w — Technical data according to DIN 12876

ParameterValue
Order-No.5507.0001.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at 20°C15 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
UN-numberUN 2857
max. delivery100 l/min
max. delivery pressure3,5 bar
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
Volume of expansion30 l
Overall dimensions WxDxH **730x804x1738 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryII
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.1.0/26

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure of at least 1 bar between inlet and outlet to operate correctly.

Ensures the cooling water circuit can provide adequate flow; insufficient differential pressure will impair cooling performance.

Source
“min. cooling water differential pressure”
Installation

The maximum permissible cooling water pressure is 6 bar, so the supply must be regulated or protected against overpressure.

Prevents damage to the water-cooled refrigeration circuit and connected components.

Source
“max. cooling water pressure”
Operation

The unit is designed for cooling water intake between 5°C and 20°C; lower temperatures require insulating the cooling water pipes inside the unit.

Operating outside this range risks condensation and requires additional insulation measures.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Sizing must account for reduced capacity and higher water demand at elevated cooling water temperatures.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Incompatible water chemistry can corrode or degrade wetted materials and cause leaks or fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%; combinations such as -5% voltage with +2% frequency are not allowed.

Defines acceptable supply quality for reliable operation and warranty compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so the customer must provide the final connection.

Affects electrical installation planning and the need for a suitable disconnector or plug on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The refrigerant is R-744 (CO2), classified A1 with GHS hazard H280, and has a GWP of 1.

Informs safety measures for pressurized gas, ventilation requirements and environmental reporting.

Source
“R-744 (A1, H280)”
Installation

The unit has protection degree IP20, so it must be installed in a dry indoor environment protected from water ingress.

Prevents electric shock and equipment damage; IP20 is not suitable for wet or outdoor locations.

Source
“Degree of Protection”
Operation

Ambient temperature must remain between 5°C and 40°C for specified operation.

Operating outside this range may reduce performance or damage the refrigeration system.

Source
“min. ambient temperature”
Installation

The pressure equipment category is II, which determines the applicable pressure vessel compliance requirements.

Affects documentation, marking and conformity assessment obligations for the installed equipment.

Source
“Pressure equipment category”
Conformité

The equipment is classified for electromagnetic disturbance as EN55011 Class A, Group 1.

Class A equipment is intended for industrial environments and may require EMC measures in residential areas.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”