Unichiller GL 200w Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 1 table, 27 rows

Unichiller GL 200w — Technical data according to DIN 12876

ParameterValue
Order-No.5508.0004.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at 20°C19 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
UN-numberUN 2857
max. delivery100 l/min
max. delivery pressure3,5 bar
Pump connectionG1 1/4 male
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
Volume of expansion30 l
Overall dimensions WxDxH **730x804x1738 mm
Power supply (3 Phase)460V 3~ 60Hz
Pressure equipment categoryII
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.1.0/26

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires a three-phase 460 V 60 Hz supply, so the site must provide this voltage and frequency.

Determines electrical infrastructure and whether a transformer or different model is needed.

Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation

The cooling water circuit needs a differential pressure of at least 1 bar and must not exceed 6 bar.

Ensures the water-cooled condenser receives adequate flow without overpressurizing the unit.

Source
“min. cooling water differential pressure 1 bar”
Installation

Cooling water connections are G3/4 male, so mating hoses and fittings must match this thread.

Prevents delays and leaks during commissioning due to incompatible fittings.

Source
“Cooling water connection G3/4 male”
Operation

The unit is designed for cooling water intake between 5 °C and 20 °C; below the condensation point the cooling water pipes inside the unit must be insulated.

Operating outside this range risks condensation, corrosion, or reduced performance.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Rated cooling power of 19 kW applies only at 20 °C room temperature, 15 °C cooling water inlet, and 1 bar differential pressure.

Actual capacity will differ if site conditions deviate from these reference values.

Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Prevents selecting a unit for a supply that falls outside the permitted combined tolerance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Conformité

External branch circuit protection according to UL 489 is required.

Affects electrical panel design and compliance with North American installation codes.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Three-phase devices drawing less than 63 A are supplied with cable but without a plug.

The installer must provide a suitable plug or hard-wired connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The refrigerant is R-744 (A1, H280), which is a non-flammable but high-pressure gas with an asphyxiation hazard in enclosed spaces.

Informs ventilation and safety measures for the machine room.

Source
“R-744 (A1, H280)”
Selection

The chiller is water-cooled and uses natural refrigerant R-744 with GWP 1.

Requires a cooling water supply and may satisfy low-GWP refrigerant preferences.

Source
“water-cooled, natural refrigerant”
Installation

The unit has IP20 protection, so it must be installed in a dry, indoor environment protected from water ingress.

Prevents placing the chiller where splashing or condensation could cause electrical faults.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must remain between 5 °C and 40 °C for safe operation.

Ensures the chiller is not installed in a location that exceeds its operating envelope.

Source
“min. ambient temperature 5 °C”