Unichiller P007w-H Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and key features of the chiller
  2. Spec p. 2
    Technical data according to DIN 12876
  3. p. 2 table, 15 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow -20°C36 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **350x496x622 mm
Net weight57 kg
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current14 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S586152

** Please respect space requirements. See operating conditions at www.huber-online.com

Unichiller P007w-H — Technical data according to DIN 12876

ParameterValue
Order-No.3012.0265.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at ambient temperature +20°C at 15°C0,7 kW
Cooling power at ambient temperature +20°C at 0°C0,55 kW
Cooling power at ambient temperature +20°C at -10°C0,4 kW
Cooling power at ambient temperature +20°C at -20°C0,2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,2 kg
CO2 equivalent0,3 t
Circulation pumpB
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C66 l/h
Consumption at water 15°C, flow 0°C54 l/h

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires cooling water inlet temperature between 5°C and 20°C, and below the condensation point the cooling water pipes inside the unit must be insulated.

Defines the permissible cooling water supply temperature range and insulation requirement for correct installation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

The cooling water circuit must be supplied with at least 1 bar differential pressure and must never exceed 6 bar.

Determines the required cooling water supply pressure and the maximum allowable pressure for safe connection.

Source
“min. cooling water differential pressure”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides selection of cooling water quality and avoids corrosion or damage to wetted materials.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant plastic pump ordered.

Prevents chemical attack on the pump when selecting the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Specifies the recommended antifreeze approach for low-temperature operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Conformité

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed.

Defines the acceptable mains supply quality for reliable operation and warranty compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

Ensures the correct mains plug is ordered for the installation site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Selection

The unit is classified as EN55011 Class A Group 1 for electromagnetic disturbance, which affects its suitability in residential environments.

Class A equipment may not be suitable for residential use and must be considered during site selection.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection

The refrigeration machine is water-cooled and uses CFC- and HCFC-free refrigerant R-449A with GWP 1397 and 0.2 kg charge.

Informs refrigerant type, environmental impact and the need for a cooling water supply.

Source
“Refrigeration machine water-cooled, CFC- and HCFC-free”
Performance

Cooling power drops from 0.7 kW at +20°C ambient to 0.2 kW at -20°C, so low-temperature duty must be sized accordingly.

Allows correct sizing of cooling capacity across the operating temperature range.

Source
“Cooling power at ambient temperature +20°C”
Installation

The unit requires a 208-240V 1~/2~ 50/60Hz supply with maximum current 14 A and a 16 A fuse.

Defines electrical supply and protection requirements for installation.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz”
Installation

The unit has IP20 protection and must be operated between 5°C and 40°C ambient temperature.

Sets environmental limits and ingress protection for the installation location.

Source
“Degree of Protection IP20”
Maintenance

The unit has a minimum filling capacity of 3.8 l and an expansion volume of 1.7 l, which must be respected when filling the bath.

Ensures correct fluid fill level and expansion allowance for safe operation.

Source
“min. filling capacity 3,8 l”
Selection

The pump delivers 25 l/min maximum and 2.5 bar maximum pressure, with delivery falling to 6 l/min at 2.0 bar.

Allows matching pump performance to the pressure drop of the external application.

Source
“max. delivery 25 l/min”