Induztr
Résumé technique Visionneuse de documents Modèles associés

Unichiller P007w

Sommaire

  1. Overview p. 1
    Introduces the Unichiller P007w chiller and its main features
  2. Spec p. 2
    Lists technical specifications according to DIN 12876
  3. p. 2 table, 13 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **350x496x622 mm
Net weight56 kg
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current6 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S603640

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller P007w (Order-No. 3012.0217.01)

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 15°C0,7 kW
at 0°C0,55 kW
at -10°C0,4 kW
at -20°C0,2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,18 kg
CO2 equivalent0,3 t
Circulation pumpB
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 15°C66 l/h
Consumption at water 15°C, flow 0°C54 l/h
Consumption at water 15°C, flow -20°C36 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar

Строка «Cooling power at ambient temperature +20°C» напечатана как заголовок группы; значения kW относятся к подстрокам at 15°C / 0°C / -10°C / -20°C.

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.

To ensure the cooling water circuit can provide adequate flow and pressure without exceeding component limits.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit is designed for cooling water intake temperatures between 5°C and 20°C, and if the water temperature falls below the condensation point, the cooling water pipes inside the unit must be insulated.

To avoid condensation and ensure proper operation within the specified cooling water temperature range.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Operation

The plastic pump is not resistant to acetone and polyglycols, so these fluids must not be used; instead, water mixed with glysantine or ethylene glycol is recommended for freeze protection.

To prevent pump damage and ensure compatibility of the heat transfer fluid with the wetted materials.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water should be used to avoid corrosion or fouling.

To select appropriate cooling water quality and avoid damage to the cooling circuit materials.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité

Voltage tolerance is ±5% with a simultaneous frequency tolerance of ±2%, but combinations such as -5% voltage and +2% frequency are not allowed.

To ensure the power supply remains within safe operating limits and avoid invalid tolerance combinations.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Single-phase and two-phase devices are supplied with a power cable and country-specific plug, which must be specified when ordering.

To ensure the correct power connection is supplied for the installation site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The refrigerant R-449A has a Global Warming Potential of 1397 and a CO2 equivalent of 0.3 t, which must be considered for environmental compliance and leak management.

To assess environmental impact and comply with refrigerant regulations.

Source
“Global Warming Potential (GWP) 1397”
Performance

Cooling power drops as the cooling water temperature increases, and the cooling water flow rate may need to be increased accordingly.

To properly size the cooling water supply and predict performance under varying conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The unit has a degree of protection IP20, so it must be installed in a dry environment protected from water ingress.

To ensure the installation location meets the environmental protection requirements.

Source
“Degree of Protection IP20”
Operation

The minimum ambient temperature is 5°C and the maximum is 40°C, so the unit must be operated within this range.

To avoid operating outside the specified ambient conditions that could affect performance or safety.

Source
“min. ambient temperature 5 °C”
Selection

The circulation pump is a high-pressure type (model B) with a maximum delivery pressure of 2.5 bar, suitable for applications with high pressure drops.

To match the pump capability to the pressure drop of the external circuit.

Source
“Unichiller „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops.”
Maintenance

The unit has a level indicator with sight glass, allowing visual monitoring of the fill level.

To facilitate routine checks of the heat transfer fluid level.

Source
“Level indicator with sight glass.”