Selection
The chiller is rated for an operating temperature range of -20 to 100 °C, so the process must stay within these limits.
Defines the application window and excludes processes requiring lower or higher temperatures.
Source
“Operating temperature range
-20...100 °C”
Performance
Cooling capacity falls sharply at low temperature, from 1 kW at 15 °C to only 0.15 kW at -20 °C, so the load must be checked at the coldest set point.
Prevents undersizing when the process requires deep cooling.
Source
“at -20°C
0,15 kW”
Installation
The unit requires a 220-240 V single- or two-phase supply with a maximum current of 14 A and a 16 A fuse.
Determines electrical supply, breaker rating and cabling before installation.
Source
“Power supply requirement
220-240V 1~/2~ 50/60Hz
max. current
14 A
Fuse
16 A”
Installation
The chiller is air-cooled and must be installed where ambient temperature stays between 5 °C and 40 °C.
Ambient conditions affect cooling performance and reliable operation.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Performance
Cooling capacity is specified at 20 °C room temperature and may drop if ambient temperature rises.
Prevents overestimating available cooling in warm plant rooms.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Safety
The refrigerant is R-449A with GHS classification H280 and a GWP of 1397, so refrigerant handling and leak response must follow safety rules.
Affects safety planning, ventilation and refrigerant leak procedures.
Source
“R-449A (A1, H280)”
Safety
The plastic pump is not resistant to acetone or polyglycols, so these fluids must not be used without a more resistant pump option.
Prevents chemical attack and premature pump failure in fluid selection.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection the manufacturer recommends mixing water with glysantine or ethylene glycol rather than using pure water.
Guides heat-transfer fluid selection for sub-zero operation.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
The pump connection is G3/4 male, so hoses and fittings must match this thread.
Ensures correct plumbing connections during installation.
Source
“Pump connection
G3/4 male”
Performance
The circulation pump delivers 25 l/min maximum and 2.5 bar maximum pressure, with flow dropping to 6 l/min at 2.0 bar.
Allows checking whether the pump can overcome the pressure drop of the external circuit.
Source
“max. delivery
25 l/min
max. delivery pressure
2,5 bar”
Installation
The unit needs at least 3.8 l filling capacity and has 1.7 l expansion volume, which sets the minimum fluid charge and expansion allowance.
Determines fluid volume and expansion vessel sizing for the loop.
Source
“min. filling capacity
3,8 l
Volume of expansion
1,7 l”
Conformité
The equipment is classified as pressure equipment category Art. 4.3 PED and has IP20 protection, so it must be installed in a dry indoor location.
Affects compliance documentation and environmental protection of the installation.
Source
“Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20”
Installation
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and combining -5% voltage with +2% frequency is not allowed.
Prevents operation on unstable or mismatched power supplies.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at ordering.
Ensures the correct plug is ordered for the installation site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”