Unichiller P010 OLÉ Technical Data

Sommaire

  1. Overview p. 1
    Product name and cover page of the datasheet
  2. Spec p. 2
    Technical data according to DIN 12876

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 40 °C, so it must be matched to the process temperature window.

Defines whether the unit can cover the required process setpoints before selection.

Source
“Operating temperature range -20...40 °C”
Performance

Cooling capacity falls from 1 kW at 15 °C to 0.15 kW at -20 °C, so low-temperature duty needs a larger unit or lower heat load.

Cooling capacity must be checked at the actual operating temperature, not just at nominal conditions.

Source
“at 15°C 1 kW at 0°C 0,8 kW at -10°C 0,5 kW at -20°C 0,15 kW”
Installation

The refrigeration machine is air-cooled, so adequate room ventilation and ambient temperature control are required at the installation site.

Air-cooled units reject heat to the room and need sufficient airflow and ambient conditions.

Source
“Refrigeration machine air-cooled, CFC- and HCFC-free”
Safety

The refrigerant is R-449A with GHS classification H280, so leak detection and ventilation provisions for a pressurized gas are needed.

Refrigerant safety class and hazard statement affect site safety planning and handling procedures.

Source
“R-449A (A1, H280)”
Conformité

The refrigerant charge corresponds to 0.4 t CO2 equivalent, which may trigger F-gas reporting or leak-check obligations.

CO2 equivalent determines regulatory leak-check and documentation requirements in many jurisdictions.

Source
“CO2 equivalent 0,4 t”
Installation

The unit requires a 110-120 V single-phase 50/60 Hz supply with 13 A max current and a 16 A fuse, so the electrical circuit must be sized accordingly.

Electrical supply and protection rating must be confirmed before installation.

Source
“Power supply requirement 110-120V 1~ 50/60Hz max. current 13 A Fuse 16 A”
Installation

The pump connection is G3/4 male, so hoses and fittings must match this thread size.

Thread mismatch prevents connection and may require adapters.

Source
“Pump connection G3/4 male”
Operation

The pump delivers 21 l/min at 0.5 bar but only 6 l/min at 2.0 bar, so high-pressure-drop circuits reduce available flow.

Flow must be verified against the pressure drop of the external circuit to avoid undercooling.

Source
“at 0,5 bar 21 l/min at 1,0 bar 17 l/min at 1,5 bar 11 l/min at 2,0 bar 6 l/min”
Operation

The minimum filling capacity is 3.8 l and the expansion tank is 1.7 l, so the loop volume and thermal expansion must be checked.

Insufficient loop volume or expansion capacity can cause poor temperature control or overflow.

Source
“min. filling capacity 3,8 l expansion tank 1,7 l”
Installation

The unit has IP20 protection and must be installed in an environment between 5 and 40 °C ambient temperature.

IP20 offers no water protection and ambient limits constrain the installation location.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity data are valid at 20 °C room temperature and may drop if ambient temperature rises.

Actual available cooling capacity depends on the installation room temperature.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation

The plastic pump is not resistant to acetone and polyglycols, so these fluids must be avoided or a more resistant pump specified.

Fluid compatibility determines pump life and must be checked before selecting the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Prevents freezing and damage in the cooling circuit at low temperatures.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and combined worst-case deviations are not allowed.

Supply quality must be verified to avoid invalid operating conditions and equipment damage.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”