Product introduction and model identification for Unichiller P012
Specp. 2
Performance and technical specifications according to DIN 12876
p. 1table, 32 rows
Unichiller P012 — Technical data according to DIN 12876
Parameter
Value
Order-No.
3009.0161.01
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at ambient temperature +20°C
at 15°C
1,2 kW
at 0°C
1 kW
at -10°C
0,7 kW
at -20°C
0,25 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,35 kg
CO2 equivalent
0,5 t
Circulation pump:
B
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
21 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 1,5 bar
11 l/min
Delivery at 2,0 bar
6 l/min
Pump connection
G3/4 male
min. filling capacity
3,8 l
Volume of expansion
1,7 l
Overall dimensions WxDxH **
420x487x579 mm
Net weight
65 kg
sound pressure level +/- 4 dB(A)
59 dB(A)
Power supply requirement
208V 2~ 60Hz
** — сноска в исходном тексте не раскрыта.
Engineering Facts
14 facts · every claim linked to a page
Selection
The chiller is rated for an operating temperature range of -20 to 40 °C, so the process must stay within these limits.
Defines the application window and excludes processes requiring lower or higher temperatures.
Source
“Operating temperature range
-20...40 °C”
Performance
Cooling capacity drops sharply at low temperatures, from 1.2 kW at +15 °C to only 0.25 kW at -20 °C, so sizing must use the capacity at the actual set point.
Prevents undersizing when the process runs cold; capacity at -20 °C is roughly one-fifth of that at +15 °C.
Source
“at 15°C
1,2 kW
at 0°C
1 kW
at -10°C
0,7 kW
at -20°C
0,25 kW”
Performance
Cooling capacity is specified at 20 °C room temperature and may fall if ambient rises, so the installation room must be kept cool.
Air-cooled units lose capacity in warm rooms; ambient conditions must be considered during selection and site planning.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit requires a 208 V two-phase 60 Hz supply with a 7 A max current and 16 A fuse, which must be provided at the installation site.
Determines the electrical supply, cabling and protection device needed before installation.
Source
“Power supply requirement
208V 2~ 60Hz”
Installation
Voltage and frequency tolerances are ±5 % and ±2 % but must not be combined in the unfavorable direction, so supply quality must be checked.
Avoids operating the compressor outside allowed supply conditions, which could damage it or void warranty.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices below 63 A are delivered with cable but without plug, so a suitable plug must be sourced and fitted on site.
Affects electrical installation planning and the bill of materials for commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The refrigerant R-449A is classified A1 with hazard statement H280 (pressurized gas), so handling and ventilation precautions apply.
Informs safety assessment, labeling and room ventilation requirements for the refrigerant charge.
Source
“R-449A (A1, H280)”
Conformité
The refrigerant has a GWP of 1397 and a CO2 equivalent of 0.5 t, which matters for F-gas reporting and leak-check obligations.
Determines regulatory obligations for refrigerant containment, leak checks and documentation.
Source
“Global Warming Potential (GWP)
1397”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must not be used as heat transfer fluid.
Prevents chemical attack and pump failure; an alternative resistant plastic is only available on request.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Avoids freezing and damage when operating near or below 0 °C.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
The pump connection is G3/4 male and hose couplings are included, so mating hoses and fittings must match this thread.
Ensures correct plumbing interface and avoids adapters or leaks at commissioning.
Source
“Pump connection
G3/4 male”
Installation
The unit needs a minimum filling capacity of 3.8 l plus 1.7 l expansion volume, so the external loop volume must be planned accordingly.
Determines heat transfer fluid quantity and expansion vessel sizing for the installation.
Source
“min. filling capacity
3,8 l”
Installation
The unit is IP20 and must be installed where ambient temperature stays between 5 and 40 °C.
IP20 offers no water protection and the ambient limits define acceptable site conditions.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Maintenance
The unit emits 59 dB(A) sound pressure, which must be considered for workplace noise and room placement.
Affects installation location and possible hearing protection or acoustic measures.