Induztr
Résumé technique Visionneuse de documents Modèles associés

Unichiller P015

Sommaire

  1. Overview p. 1
    Product introduction and main features of Unichiller P015
  2. Spec pp. 1-2
    Technical data, performance and dimensions per DIN 12876
  3. p. 2 table, 7 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
max. current8,5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S609077

Unichiller P015 — Technical data according to DIN 12876

ParameterValue
Order-No.3051.0046.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 15°C1,5 kW
at 0°C1 kW
at -10°C0,7 kW
at -20°C0,3 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,16 kg
CO2 equivalent0,0 t
Circulation pump:B
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **420x487x579 mm
Net weight61 kg
sound pressure level +/- 4 dB(A)63 dB(A)
Power supply requirement220-240V 1~/2~ 50Hz

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Selection

The chiller uses R-290 (A3, H220) refrigerant, which is flammable, so the installation location must be assessed for A3 refrigerant safety requirements.

A3 flammability classification affects room ventilation, ignition-source control, and local refrigerant regulations during selection and installation.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Selection

Cooling capacity drops from 1.5 kW at +15 °C to 0.3 kW at -20 °C, so the required load must be checked at the actual operating temperature.

Sizing only on nominal capacity will undersize the unit at low temperatures and cause insufficient cooling.

Source
“at 15°C 1,5 kW at 0°C 1 kW at -10°C 0,7 kW at -20°C 0,3 kW”
Performance

Cooling capacity is rated at 20 °C room temperature and may fall if ambient temperature rises, so the installation room must stay within the ambient limits.

High ambient temperature reduces available cooling and can make the chiller unable to hold the set point.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit requires 220-240 V single/two-phase 50 Hz supply with 8.5 A max current and a 16 A fuse, so the electrical supply and protection must be sized accordingly.

Correct breaker/fuse rating and supply configuration are needed to avoid nuisance trips and ensure safe operation.

Source
“max. current 8,5 A Fuse 16 A”
Installation

Voltage and frequency tolerances are ±5% and ±2% but must not be combined in the unfavorable direction, so supply quality must be verified before commissioning.

Out-of-tolerance supply can damage the compressor and pump and void warranty.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The pump connection is G3/4 male and the minimum filling capacity is 3.8 l, so hoses and expansion volume must be planned for this circuit.

Mismatched fittings or insufficient fill volume will prevent proper commissioning and circulation.

Source
“Pump connection G3/4 male min. filling capacity 3,8 l”
Operation

The plastic pump is not resistant to acetone and polyglycols, so these fluids must not be used without a more resistant pump option.

Using incompatible fluid will damage the pump and contaminate the loop.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Prevents freezing and damage when operating near or below 0 °C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The unit is IP20 and must be installed in an environment between 5 °C and 40 °C ambient temperature.

IP20 offers no water protection and ambient limits define where the chiller can be safely located.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Conformité

The equipment is EMC Class A Group 1, so it may cause radio interference in residential areas and is intended for industrial environments.

Class A equipment may not be suitable for residential use without additional EMC measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are supplied with a country-specific plug that must be specified at order, while three-phase devices below 63 A come without a plug.

Missing plug specification at order leads to delays and additional electrical work on site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Maintenance

The refrigerant charge is only 0.16 kg with GWP 0.02, so leak-check and service procedures must account for a small A3 charge.

Small flammable charge affects service safety and refrigerant record-keeping obligations.

Source
“Refrigerant quantity 0,16 kg CO2 equivalent 0,0 t”
Performance

The pump delivers 25 l/min max and 6 l/min at 2.0 bar, so system pressure drop must be checked against the pump curve.

High-pressure-drop loops may need the P model's higher discharge pressure or a different pump selection.

Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar”