Performance, temperature, and dimensional data per DIN 12876
p. 2table, 12 rows
p. 1table, 32 rows
Technical data according to DIN 12876
Parameter
Value
min. filling capacity
18 l
Volume of expansion
48 l
Overall dimensions WxDxH **
740x1160x1165 mm
Net weight
299 kg
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
15 A
Fuse (3 phase)
3x16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S625743
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Unichiller P100-H
Order-No.
3059.0004.01
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at ambient temperature +20°C
at 100°C
10 kW
at 20°C
10 kW
at 0°C
7,5 kW
at -10°C
4,7 kW
at -20°C
2,4 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
1,1 kg
CO2 equivalent
1,5 t
Circulation pump:
E2
max. delivery
130 l/min
max. delivery pressure
5,7 bar
Delivery at 2,4 bar
116 l/min
Delivery at 2,5 bar
115 l/min
Delivery at 3,0 bar
108 l/min
Delivery at 3,5 bar
100 l/min
Delivery at 4,0 bar
92 l/min
Delivery at 4,5 bar
80 l/min
Delivery at 5,0 bar
59 l/min
Delivery at 5,5 bar
20 l/min
Pump connection
G1 1/4 male
Engineering Facts
14 facts · every claim linked to a page
Selection
The chiller is intended only for externally closed applications, so an open-loop or vented external circuit is not acceptable.
Determines whether the unit can be applied to the customer's process circuit at all.
Source
“For externally closed applications.”
Selection
The P-series pumps are specified for high pressure-drop applications, so they suit restrictive external circuits better than standard pumps.
Guides pump/model choice when the external loop has significant pressure loss.
Source
“Unichiller „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops.”
Performance
Cooling capacity is only valid at 20 °C room temperature and drops as ambient temperature rises, so capacity must be derated for warmer rooms.
Prevents overestimating available cooling at elevated ambient conditions.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The three-phase unit is supplied with a cable but without a plug, so the installer must provide a suitable 3-phase plug/connection.
Affects electrical installation scope and materials before commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances are linked: ±5 % voltage is allowed only with simultaneous ±2 % frequency, and some combinations are explicitly not allowed.
Needed to verify site power quality before connecting the chiller.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit requires 3-phase 400 V 50 Hz supply with max current 15 A and 3x16 A fusing.
Defines the electrical supply and protection rating for installation planning.
Source
“max. current (3 Phase)
15 A”
Installation
The chiller must be installed where ambient temperature stays between 5 °C and 40 °C.
Sets the allowable room/installation environment for reliable operation.
Source
“min. ambient temperature
5 °C”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions apply.
Informs safety assessment and refrigerant handling requirements on site.
Source
“R-449A (A1, H280)”
Conformité
The unit is EMC Class A Group 1, which is industrial-grade emission classification and may not be suitable for residential environments.
Affects EMC compatibility assessment and installation location restrictions.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The unit has IP20 protection, so it must be installed in a dry, protected location and not exposed to water or dust ingress.
Constrains the installation environment and enclosure requirements.
Source
“Degree of Protection
IP20”
Selection
The unit contains 1.1 kg of R-449A with GWP 1397, equivalent to 1.5 t CO2, which matters for refrigerant reporting and leak obligations.
Needed for environmental compliance and refrigerant charge documentation.
Source
“CO2 equivalent
1,5 t”
Operation
The controller supports Ethernet, USB host/device and RS232, enabling remote control and data transmission via PC or network.
Determines available integration options for monitoring and remote control.
Source
“Interface digital
Ethernet, USB (Host u.
Device), RS232”
Maintenance
The expansion tank is atmospheric open and has a digital level indicator, so level monitoring is part of routine operation.
Indicates a maintenance/monitoring point for the heat transfer fluid level.
Source
“With digital level indicator.”
Installation
Space requirements around the unit must be respected per the operating conditions, so clearance planning is mandatory.
Prevents installation in tight spaces that would impair operation or service access.
Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”