Product introduction and key features of the chiller
Specpp. 1-2
Technical data and performance specifications per DIN 12876
p. 2table, 7 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Overall dimensions WxDxH **
940x1290x1130 mm
Power supply (3 Phase)
400V 3~ 50Hz
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/21
** Please respect space requirements. See operating conditions at www.huber-online.com
Unichiller P180w — Technical data according to DIN 12876
Parameter
Value
Order-No.
3041.0003.01
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at ambient temperature +20°C
at 20°C
18 kW
at 0°C
10 kW
at -10°C
6 kW
at -20°C
3,5 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
1,49 kg
CO2 equivalent
2,1 t
Circulation pump
E2
max. delivery
130 l/min
max. delivery pressure
5,7 bar
Delivery at 2,4 bar
116 l/min
Delivery at 2,5 bar
115 l/min
Delivery at 3,0 bar
108 l/min
Delivery at 3,5 bar
100 l/min
Delivery at 4,0 bar
92 l/min
Delivery at 4,5 bar
80 l/min
Delivery at 5,0 bar
59 l/min
Delivery at 5,5 bar
20 l/min
Pump connection
G1 1/4 male
Cooling water connection
G3/4 male
min. filling capacity
18 l
Volume of expansion
48 l
Order-No. 3041.0003.01; cooling power values are listed under the heading 'Cooling power at ambient temperature +20°C'.
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires cooling water inlet temperature between 5°C and 20°C, so the site must provide a suitable cooling water supply.
Cooling water availability and temperature determine feasibility of installation and required utilities.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water temperature can fall below the condensation point, the cooling water pipes inside the unit must be insulated.
Prevents condensation and corrosion/electrical risk in the cooling water circuit.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling capacity drops as cooling water temperature rises, and cooling water flow rate may need to increase.
Affects sizing of cooling water supply and expected performance at site conditions.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water should be used.
Guides water quality and material compatibility to avoid corrosion and fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Conformité
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Ensures the electrical supply meets EN60034-1 tolerances for reliable operation.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices with current consumption less than 63A are delivered with cable but without plug.
Installer must plan for a suitable plug or direct connection on site.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Three-phase devices with current consumption greater than 63A are delivered without cable and without plug.
Requires on-site power cabling and connection by qualified personnel.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation must be considered.
Informs safety measures for refrigerant leak and service procedures.
Source
“R-449A (A1, H280)”
Selection
The unit has a CO2 equivalent of 2.1 t based on 1.49 kg of R-449A with GWP 1397, relevant for refrigerant regulations.
Needed for environmental compliance and leak-check obligations.
Source
“CO2 equivalent 2,1 t”
Operation
The pump delivers 130 l/min maximum and 5.7 bar maximum delivery pressure, with flow dropping at higher pressure.
Determines whether the pump can overcome system pressure drop in external closed loops.
Source
“max. delivery 130 l/min”
Installation
Pump connection is G1 1/4 male and cooling water connection is G3/4 male, so matching hoses and fittings are required.
Ensures correct plumbing adapters are available before installation.
Source
“Pump connection G1 1/4 male”
Selection
Minimum filling capacity is 18 l and expansion volume is 48 l, which must be considered for external loop volume.
Affects heat transfer fluid quantity and expansion tank sizing.
Source
“min. filling capacity 18 l”
Installation
The unit requires a 400V 3~ 50Hz three-phase power supply and has IP20 protection, so it must be installed in a dry indoor location.
Defines electrical supply and environmental protection requirements for siting.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Operation
Ambient temperature must be between 5°C and 40°C for operation.
Ensures the installation room stays within allowable ambient conditions.