Induztr
Résumé technique Visionneuse de documents Modèles associés

Unichiller P230-H Datasheet

Sommaire

  1. Overview p. 1
    Introduces the Unichiller P230-H chiller and its main features
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 7 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
Overall dimensions WxDxH **938x1288x2003 mm
Power supply (3 Phase)460V 3~ 60Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Unichiller P230-H — Technical data according to DIN 12876

ParameterValue
Order-No.3039.0024.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at ambient temperature +20°C
at 100°C23 kW
at 20°C23 kW
at 0°C13,5 kW
at -10°C9 kW
at -20°C5,5 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,49 kg
CO2 equivalent2,1 t
Circulation pump:E2
max. delivery140 l/min
max. delivery pressure5,3 bar
Delivery at 2,8 bar126 l/min
Delivery at 3,0 bar124 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar103 l/min
Delivery at 4,5 bar83 l/min
Delivery at 5,0 bar43 l/min
Pump connectionG1 1/4 male
min. filling capacity18 l
Volume of expansion48 l

Строка «Cooling power at ambient temperature +20°C» напечатана как заголовок группы; значения при разных температурах приведены ниже.

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 100 °C, so it must be matched to the process temperature window.

Defines whether the unit can cover the required process temperatures before selection.

Source
“Operating temperature range -20...100 °C”
Performance

Cooling capacity falls from 23 kW at 20 °C to 5.5 kW at -20 °C, so sizing must use the capacity at the actual operating temperature.

Prevents oversizing or undersizing when the process runs at low temperature.

Source
“at 20°C 23 kW at 0°C 13,5 kW at -10°C 9 kW at -20°C 5,5 kW”
Installation

The unit requires a 460 V three-phase 60 Hz supply, which must be confirmed at the installation site.

Electrical supply mismatch will prevent commissioning.

Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation

Voltage and frequency tolerances are linked: -5% voltage with +2% frequency is not allowed, so supply quality must be checked.

Avoids operating the compressor and pump outside the permitted supply envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Conformité

External branch circuit protection according to UL 489 is required, so the panel must include a suitable UL 489 breaker.

Needed for a compliant and safe electrical installation.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Three-phase devices below 63 A are delivered with cable but without plug, so the site must provide the connection means.

Affects planning of the electrical connection and required hardware.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.

Determines the acceptable environment and enclosure requirements at the installation site.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop above 20 °C.

Room conditions directly affect available cooling performance and must be considered in layout.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit is air-cooled and requires respect of space requirements, so clearance for air flow must be planned.

Insufficient clearance reduces cooling performance and can cause overheating.

Source
“Please respect space requirements.”
Safety

The refrigerant is R-449A with GHS hazard H280, so leak detection and ventilation provisions are needed.

Affects site safety assessment and handling procedures for the refrigerant.

Source
“R-449A (A1, H280)”
Conformité

The refrigerant charge corresponds to 2.1 t CO2 equivalent, which may trigger refrigerant regulations and record keeping.

Determines whether F-gas or equivalent reporting and leak-check obligations apply.

Source
“CO2 equivalent 2,1 t”
Installation

The pump connection is G1 1/4 male, so mating hoses and couplings must match this thread.

Prevents mismatch of process piping and accessories during installation.

Source
“Pump connection G1 1/4 male”
Operation

The minimum filling capacity is 18 l and expansion volume is 48 l, so the external loop volume must be considered.

Ensures the system has adequate heat transfer fluid and expansion space.

Source
“min. filling capacity 18 l Volume of expansion 48 l”
Selection

The unit is intended for externally closed applications, so the process loop must be closed.

Using it on an open loop would be outside the intended application.

Source
“For externally closed applications.”