Introduces the Unimotive 26w thermostatic bath/circulator unit
Specpp. 1-2
Operating limits, temperature ranges, and performance data per DIN 12876
Specp. 2
Physical dimensions, fittings, and connection details
p. 2table, 25 rows
p. 1table, 27 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 1,0 bar
179 l/min
Delivery at 2,0 bar
154 l/min
Delivery at 3,0 bar
124 l/min
Delivery at 4,0 bar
87 l/min
Delivery at 5,0 bar
42 l/min
Pump connection
M38x1,5 male
max. permissible kin. viscosity
200 mm²/s
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 0°C
1260 l/h
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
6 bar
min. filling capacity
8,9 l
Filling capacity expansion tank
19,9 l
Overall dimensions WxDxH **
730x804x1738 mm
Net weight
462 kg
Power supply requirement
400V 3~ 50Hz
max. current
38 A
Fuse
3x40A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S551405
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Order-No.
5005.0001.01
Model
Unimotive 26w
Operating temperature range
-45...95 °C
Temperature stability at -10°C
0,05 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
24 kW
Cooling power with Water/ Glysantin 40/60 at 20°C
28 kW
Cooling power with Water/ Glysantin 40/60 at 0°C
25 kW
at -20°C
14,5 kW
at -30°C
6,5 kW
at -40°C
2,6 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
3,6 kg
CO2 equivalent
7,7 t
Circulation pump: MK pump max. delivery
201 l/min
max. delivery pressure
5,3 bar
Engineering Facts
13 facts · every claim linked to a page
Selection
The unit is intended for water-ethylene glycol mixtures with corrosion protection down to -45°C, so it suits automotive temperature simulation and material testing rather than general lab use.
Confirms the heat transfer fluid and application envelope before specifying the unit.
Source
“The temperature control systems are constructed for operation with Water-ethylene glycol mixture incl. corrosion protection (for example Glysantin®) down to -45°C.”
Installation
The cooling water supply must be between 5°C and 20°C, and pipes must be insulated if the water temperature can fall below the dew point.
Defines the site cooling water requirements and condensation risk during installation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
Cooling water differential pressure must stay between 1 bar and 4 bar, with a maximum supply pressure of 6 bar.
Ensures the facility cooling loop can provide adequate flow without overpressurizing the unit.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit requires a 400 V 3~ 50 Hz supply with a maximum current of 38 A and 3x40 A fusing.
Determines electrical supply, cable sizing, and protection device selection.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must provide the plug and connection.
Affects electrical installation planning and who supplies the plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
Leakage current exceeds 3.5 mA, which may require additional protective measures such as a higher-rated RCD or fixed connection.
Critical for electrical safety design and RCD selection.
Source
“Attention: leakage current > 3,5mA”
Operation
Cooling capacity drops significantly at lower temperatures, from 28 kW at 20°C to 2.6 kW at -40°C, so low-temperature loads must be sized accordingly.
Prevents oversizing expectations for deep-cooling applications.
Source
“at -40°C 2,6 kW”
Operation
The pump delivers 201 l/min at zero head but only 42 l/min at 5 bar, so system pressure drop directly limits available flow.
Needed to match pump performance to the external circuit resistance.
Source
“max. delivery 201 l/min”
Operation
Maximum permissible kinematic viscosity of the heat transfer fluid is 200 mm²/s, limiting use of very viscous fluids at low temperature.
Constrains fluid selection, especially for low-temperature operation.
Source
“max. permissible kin. viscosity 200 mm²/s”
Installation
The unit is IP20 and must be installed where ambient temperature stays between 5°C and 40°C.
Defines environmental limits and protection against dust and water ingress.
Source
“Degree of Protection IP20”
Conformité
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.
Important for sites with unstable power quality to avoid invalid operating conditions.
Source
“-5% voltage and +2% frequency -> not allowed!”
Selection
The refrigerant R-452A has a GWP of 2141 and a charge of 3.6 kg, giving a CO2 equivalent of 7.7 t, which matters for environmental reporting and leak management.
Relevant for F-gas compliance and environmental impact assessment.
Source
“Global Warming Potential (GWP) 2141”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and corrosion control to protect the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”