Unistat 430w Technical Data

Sommaire

  1. Overview p. 1
    Product introduction and general description of the Unistat 430w unit
  2. Spec pp. 1-2
    Technical data per DIN 12876 including performance and dimensions
  3. p. 2 table, 19 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C312 l/h
Consumption at water 15°C, flow 0°C312 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity4 l
Filling capacity expansion tank3,7 l
Overall dimensions WxDxH **460x554x1453 mm
Net weight175 kg
sound pressure level +/- 4 dB(A)61 dB(A)
Power supply requirement400V 3~ 50Hz
max. current14,5 A
Fuse3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S393709

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat 430w (Order-No.: 1069.0002.01)

ParameterValue
Operating temperature range-40...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power4 kW
Cooling power with Thermooil at 250°C3,5 kW
Cooling power with Thermooil at 200°C3,5 kW
Cooling power with Thermooil at 100°C3,5 kW
Cooling power with Thermooil at 20°C3,5 kW
Cooling power with Ethanol at 0°C3,5 kW
Cooling power with Ethanol at -20°C2,2 kW
Cooling power with Ethanol at -40°C0,3 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity0,97 kg
CO2 equivalent2,1 t
Circulation pump:MK pump
max. delivery91 l/min
max. delivery pressure1,5 bar
Delivery at 0,25 bar85 l/min
Delivery at 0,5 bar78 l/min
Delivery at 0,75 bar71 l/min
Delivery at 1,0 bar60 l/min
Delivery at 1,25 bar40 l/min
Pump connectionM30x1,5 male

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power distribution, cable and breaker sizing before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water circuit must be supplied at 1 bar minimum differential pressure and must not exceed 6 bar.

Ensures adequate cooling water flow and protects the water-cooled condenser from overpressure.

Source
“min. cooling water differential pressure 1 bar”
Installation

Cooling water intake temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation occurs.

Defines the permissible cooling water supply conditions and condensation protection measures.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops significantly at low temperatures with ethanol, from 3.5 kW at 0 °C to only 0.3 kW at -40 °C.

Critical for sizing the unit when low-temperature operation near the lower limit is required.

Source
“at -40°C 0,3 kW”
Performance

The circulation pump delivers 91 l/min maximum and 1.5 bar maximum pressure, with flow falling as pressure rises.

Needed to match the pump curve against the pressure drop of the external circuit.

Source
“max. delivery 91 l/min”
Safety

The unit has a leakage current above 3.5 mA, which affects residual current device selection.

Requires appropriate RCD type and rating to avoid nuisance tripping and ensure protection.

Source
“Attention: leakage current > 3,5mA”
Conformité

Voltage and frequency tolerances are linked: -5% voltage with +2% frequency is not allowed.

Prevents operation outside the EN60034-1 tolerance envelope that could damage the compressor or pump.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a plug must be provided on site.

Affects site electrical preparation and commissioning schedule.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The unit is rated for ambient temperatures from 5 °C to 40 °C and IP20 protection, so it must be installed indoors in a dry location.

Defines the allowable room conditions and excludes wet or dusty environments.

Source
“max. ambient temperature 40 °C”
Maintenance

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and treatment to prevent corrosion and fouling of the condenser.

Source
“Please use suitable cooling water.”
Selection

The refrigerant R-452A has a GWP of 2141 and a charge of 0.97 kg, giving 2.1 t CO2 equivalent.

Relevant for refrigerant regulations, leak monitoring and environmental reporting.

Source
“Global Warming Potential (GWP) 2141”
Installation

The pump connection is M30x1.5 male with a Y bore shape per DIN 3863, so matching ball-socket hoses and sleeve nuts are required.

Ensures correct mating of external tempering hoses and avoids leaks.

Source
“Pump connection M30x1,5 male”