Unistat 520w Technical Data

Sommaire

  1. Overview p. 1
    Product introduction and general description of the Unistat 520w unit
  2. Spec pp. 1-2
    Technical data and performance specifications according to DIN 12876
  3. p. 2 table, 20 rows
  4. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C492 l/h
Consumption at water 15°C, flow 0°C492 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity4,9 l
Filling capacity expansion tank16 l
Overall dimensions WxDxH **540x604x1332 mm
Net weight210 kg
sound pressure level +/- 4 dB(A)61 dB(A)
Power supply requirement400V 3~ 50Hz
max. current14 A
Fuse3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S388606

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat 520w (Order-No.: 1072.0001.01)

ParameterValue
Operating temperature range-55...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C6 kW
Cooling power with Thermooil at 200°C6 kW
Cooling power with Thermooil at 100°C6 kW
Cooling power with Thermooil at 20°C6 kW
Cooling power with Ethanol at 0°C6 kW
Cooling power with Ethanol at -20°C4,2 kW
Cooling power with Ethanol at -40°C1,5 kW
Cooling power with Ethanol at -50°C0,65 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity1,44 kg
CO2 equivalent3,1 t
Circulation pump: MK pump max. delivery79 l/min
max. delivery pressure1,5 bar
Delivery at 0,25 bar72 l/min
Delivery at 0,5 bar63 l/min
Delivery at 0,75 bar54 l/min
Delivery at 1,0 bar43 l/min
Delivery at 1,25 bar29 l/min

Значения приведены дословно; строки 'Cooling power with Thermooil' и 'Cooling power with Ethanol' в PDF напечатаны как группы с подстроками температур.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and protection device rating before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 14 A Fuse 3x16A”
Installation

The cooling water circuit needs at least 1 bar differential pressure and must not exceed 6 bar, with G1/2 male connections.

Defines the required cooling water supply pressure and connection size for proper operation.

Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation

Cooling water inlet temperature must be between 5 °C and 20 °C, and pipes must be insulated if it can drop below the dew point.

Prevents condensation and ensures rated cooling performance is available at the site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops significantly at low temperatures with ethanol, from 6 kW at 0 °C to only 0.65 kW at -50 °C.

Needed to check whether the unit can remove the process heat at the lowest required temperature.

Source
“at -50°C 0,65 kW”
Performance

The circulation pump delivers 79 l/min maximum and 43 l/min at 1.0 bar, which limits achievable flow in the external circuit.

Used to verify that the external loop pressure drop can be overcome at the required flow.

Source
“max. delivery 79 l/min max. delivery pressure 1,5 bar”
Selection

The refrigerant is R-452A with GWP 2141 and a CO2 equivalent of 3.1 t, which matters for environmental and regulatory reporting.

Affects refrigerant regulations, leak reporting obligations and environmental compliance.

Source
“Global Warming Potential (GWP) 2141”
Safety

Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for this.

Impacts electrical safety design and choice of residual current protective devices.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.

Avoids delays at commissioning because the customer must supply and fit the mains plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

Voltage and frequency tolerances are ±5% and ±2% but must not be combined in the unfavorable direction.

Ensures the site supply quality is within the allowed envelope for reliable operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit emits Class A, Group 1 disturbances per EN55011, which may require EMC measures in residential or sensitive environments.

Determines EMC compatibility and possible filtering or separation from sensitive equipment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Ambient temperature must stay between 5 °C and 40 °C and the unit needs the specified space clearances.

Affects room ventilation, layout and available performance at the installation site.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.

Guides water quality and treatment to avoid corrosion and fouling of the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Selection

The unit is rated for external closed and external open applications, but open applications need an isolation sleeve accessory.

Confirms suitability for open baths and flags the required accessory for that mode.

Source
“As well as for externally closed and also externally open applications.”
Operation

The pump connection is M30x1.5 male and hoses must use DIN 3863 ball socket and DIN 3870 sleeve nut.

Ensures correct mating fittings and leak-free connection of the external circuit.

Source
“Pump connection M30x1,5 male”