Product introduction and general description of the unit
Specpp. 1-2
Technical data and performance specifications per DIN 12876
p. 2table, 17 rows
p. 1table, 30 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 5,0 bar
42 l/min
Pump connection
M38x1,5 male
max. permissible kin. viscosity
100 mm²/s
Cooling water connection
G3/4 male
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
6 bar
min. filling capacity
8,9 l
Filling capacity expansion tank
19,9 l
Overall dimensions WxDxH **
918x963x1771 mm
Power supply requirement
400V 3~ 50Hz
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/24
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat GL P540w — Technical data according to DIN 12876
Parameter
Value
Order-No.
5040.0002.01
Operating temperature range
-50...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
12 kW
Cooling power with Thermooil at 200°C
28 kW
Cooling power with Thermooil at 100°C
28 kW
Cooling power with Thermooil at 20°C
28 kW
Cooling power with Ethanol at 0°C
25 kW
Cooling power with Ethanol at -20°C
17 kW
Cooling power with Ethanol at -40°C
6,8 kW
Cooling power with Ethanol at -50°C
4 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-744 (A1, H280)
Global Warming Potential (GWP)
1
Circulation pump: MK pump max. delivery
201 l/min
Circulation pump: MK pump max. delivery pressure
5,3 bar
Delivery at 1,0 bar
179 l/min
Delivery at 2,0 bar
154 l/min
Delivery at 3,0 bar
124 l/min
Delivery at 4,0 bar
87 l/min
Значения приведены дословно по тексту страницы; строки 'Cooling power with Thermooil' и 'Cooling power with Ethanol' объединены с соответствующими температурными точками.
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply, so the site must provide a 3~ 50 Hz feed sized for the 12 kW heater and 28 kW cooling load.
Determines electrical infrastructure and breaker/cable sizing before purchase.
Source
“Power supply requirement
400V 3~ 50Hz”
Installation
Cooling water must be supplied at 1–4 bar differential and no more than 6 bar, with 5–20 °C intake, so the facility water loop must be verified.
Insufficient or excessive cooling water pressure will prevent rated cooling performance or damage the circuit.
Source
“min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar”
Installation
Cooling water connections are G3/4 male and the pump connection is M38x1.5 male, so mating hoses and adapters must be ordered accordingly.
Prevents mismatch of process and cooling water plumbing at commissioning.
Source
“Pump connection
M38x1,5 male”
Installation
The unit is IP20, so it must be installed in a dry indoor location protected from water ingress.
Affects placement and enclosure requirements in the plant.
Source
“Degree of Protection
IP20”
Installation
Ambient temperature must stay between 5 °C and 40 °C, so the installation room needs adequate ventilation and climate control.
Operation outside this range risks derating or equipment fault.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The unit is water-cooled with natural refrigerant R-744 (GWP 1), so no refrigerant handling certification for high-GWP fluids is needed but water cooling is mandatory.
Confirms cooling medium availability requirement and low environmental impact for permitting.
Rated cooling power is stated at 20 °C room, 15 °C cooling water inlet and 1 bar differential, so actual capacity will drop if site conditions differ.
Prevents overestimating available cooling at the customer's real water temperature.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Operation
If cooling water is colder than the condensation point, the cooling water pipes inside the unit must be insulated.
Avoids condensation and corrosion inside the unit at low water temperatures.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Operation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and treatment to prevent corrosion and fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Conformité
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and combinations like -5% voltage with +2% frequency are not allowed.
Defines acceptable grid quality for warranty and reliable operation.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices under 63 A are delivered with cable but without plug, so a suitable plug must be sourced and fitted on site.
Affects electrical installation scope and cost.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The unit has adjustable overtemperature protection per DIN 12876 and safety classification III/FL, so it can be used for flammable fluids with proper setup.
Confirms suitability and required safety setting for flammable heat transfer fluids.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Performance
The MK pump delivers 201 l/min max at 5.3 bar, dropping to 42 l/min at 5.0 bar, so system pressure drop must be checked against the pump curve.
Ensures adequate flow through the external process loop.