Unistat P615 Datasheet

Sommaire

  1. Overview p. 1
    Product introduction and main features of Unistat P615
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 16 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 2,0 bar49 l/min
Delivery at 2,5 bar36 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
min. filling capacity5,65 l
Filling capacity expansion tank26,5 l
Overall dimensions WxDxH **1290x955x1537 mm
Net weight576 kg
Power supply factory configured (3 Phase)400V 3~ 50Hz
max. current (3 Phase)29 A
Fuse (3 phase)3x32A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S530550

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat P615 Technical data according to DIN 12876

ParameterValue
Order-No.1074.0008.01
Operating temperature range-60...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 200°C9,5 kW
Cooling power with Thermooil at 100°C9,5 kW
Cooling power with Thermooil at 20°C9,5 kW
Cooling power with Ethanol at 0°C9,5 kW
Cooling power with Ethanol at -20°C8 kW
Cooling power with Ethanol at -40°C4,6 kW
Cooling power with Ethanol at -60°C1,2 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity5,2 kg
CO2-Äquvivalent11,1 t
Circulation pump:MK pump
max. delivery82 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar75 l/min
Delivery at 1,0 bar68 l/min
Delivery at 1,5 bar59 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V supply with a 3x32 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and protection device rating before installation.

Source
“Power supply factory configured (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and fit the plug.

Affects electrical installation scope and procurement of the mating plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

Leakage current exceeds 3.5 mA, which may require a suitable residual current device or protective earthing arrangement.

Important for electrical safety design and RCD selection during installation.

Source
“Attention: leakage current > 3,5mA”
Installation

The unit must be operated within an ambient temperature of 5 to 40 °C, so the installation location must be kept within this range.

Site conditions outside this range may prevent proper operation or reduce performance.

Source
“min. ambient temperature 5 °C”
Performance

Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises, so capacity must be derated for warmer rooms.

Prevents undersizing the unit when installed in a warm environment.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

The pump connection is M30x1.5 male with bore shape Y (60°) per DIN 3863, so mating hoses or fittings must match this standard.

Ensures correct selection of hoses and adapters for the external circuit.

Source
“Pump connection M30x1,5 male”
Installation

The pump connection bore and mating ball socket/sleeve nut must follow DIN 3863 and DIN 3870, so non-compliant fittings should not be used.

Guarantees leak-free and mechanically compatible connections.

Source
“Bore shape Y (60°) according to DIN 3863”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, so fluid selection must respect this limit.

Using a too-viscous fluid can overload the pump and impair circulation.

Source
“max. permissible kin. viscosity 50 mm²/s”
Operation

The minimum filling capacity is 5.65 l and the expansion tank holds 26.5 l, so the system volume must be planned within these limits.

Ensures sufficient fluid volume and expansion space for safe operation.

Source
“min. filling capacity 5,65 l”
Selection

The unit uses refrigerant R-452A with GWP 2141 and a CO2 equivalent of 11.1 t, which matters for environmental compliance and leak reporting.

Affects refrigerant regulations, leak checks and environmental permitting.

Source
“R-452A (A1, H280)”
Selection

The unit is classified as pressure equipment category Art. 4.3 PED, so it falls under the Pressure Equipment Directive requirements.

Relevant for compliance documentation and acceptance at site.

Source
“Pressure equipment category Art. 4.3 PED”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.

Prevents installation in wet or dusty environments that could damage the equipment.

Source
“Degree of Protection IP20”
Installation

The overall dimensions are 1290x955x1537 mm and net weight 576 kg, so access routes and floor loading must be checked before installation.

Needed for rigging, doorway clearance and structural floor assessment.

Source
“Overall dimensions WxDxH ** 1290x955x1537 mm”
Conformité

The equipment is EMC class A group 1, so it may cause radio interference in residential areas and is intended for industrial environments.

Determines suitability for industrial versus residential installation.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”