Sommaire

  1. Overview p. 1
    Product introduction and main features of Unistat T345
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 1 table, 25 rows

Unistat T345 — Technical data according to DIN 12876

ParameterValue
Order-No.1042.0012.01
Operating temperature range65... 300 °C
Temperature stability at 70°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
External sensorPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power72 kW
Circulation pump:MK pump
max. delivery200 l/min
max. delivery pressure5,5 bar
Pump connectionM38x1,5 male
max. permissible kin. viscosity50 mm²/s
Filling capacity expansion tank56 l
Overall dimensions WxDxH **800x1007x1607 mm
Power supply (3 Phase)400V 3~ 50Hz
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.1.0/25

** Overall dimensions approximate and subject to change.

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit has no active refrigeration, so reaching the 65 °C minimum depends on application heat loss or suitable water cooling.

Prevents specifying the T345 for cooling duties below ambient without an external heat sink or water cooling.

Source
“These devices do not have an active refrigeration system.”
Installation

Three-phase devices drawing less than 63 A are supplied with cable but no plug, so the site must provide a suitable plug and connection.

Affects electrical installation planning and the bill of materials for the power connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances must be respected simultaneously; some combinations such as -5% voltage with +2% frequency are not allowed.

Guides verification of site power quality before commissioning to avoid invalid supply conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

Leakage current exceeds 3.5 mA, which matters for protective earthing and RCD selection.

Requires appropriate earth connection and possibly type-B or higher RCD coordination during installation.

Source
“Attention: leakage current > 3,5mA”
Conformité

The equipment is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.

Determines whether the unit can be installed in residential or light-industrial locations.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The pump connection is M38x1.5 male, so mating hoses and fittings must match this thread.

Ensures correct selection of connection hoses and adapters during installation.

Source
“Pump connection M38x1,5 male”
Operation

The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s, limiting fluid choice at low temperatures.

Prevents selecting a fluid too viscous for the pump at the operating temperature.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The unit requires 400 V 3~ 50 Hz three-phase supply, so site power must match this configuration.

Confirms the electrical supply needed before ordering and installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Operation

Ambient temperature must stay between 5 °C and 40 °C for safe operation.

Defines the allowable room conditions where the unit can be installed and operated.

Source
“min. ambient temperature 5 °C”
Installation

The unit has IP20 protection, so it must be kept in a dry indoor location protected from water and dust ingress.

Restricts installation to indoor, dry environments and affects enclosure planning.

Source
“Degree of Protection IP20”
Performance

The circulation pump delivers up to 200 l/min at up to 5.5 bar, which sets the hydraulic capability for the application loop.

Used to check whether the pump can overcome the pressure drop of the external circuit.

Source
“max. delivery 200 l/min”
Selection

The expansion tank holds 56 l and is not thermoregulated, which affects fluid volume and expansion allowance in closed systems.

Needed to size the external fluid volume and account for thermal expansion.

Source
“Expansion tank (not thermoregulated) for closed systems, lockable for open baths.”
Safety

The unit includes adjustable overtemperature protection according to DIN 12876, which must be set for the application.

Important for configuring a safe high-temperature limit during commissioning.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Installation

Space requirements must be respected, so clearance around the 800x1007x1607 mm unit must be planned.

Prevents installing the unit in a location without adequate access and ventilation clearance.

Source
“Please respect space requirements.”