Product name and general introduction to the Unistat T350 HT unit
Specp. 2
Technical data according to DIN 12876 standard
p. 1table, 25 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
65... 300 °C
Temperature stability at 70°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
External sensor
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
96 kW
Cooling power at 300°C
30 kW
Circulation pump:
MK pump
max. delivery
200 l/min
max. delivery pressure
5,5 bar
Pump connection
M38x1,5 male
max. permissible kin. viscosity
50 mm²/s
Filling capacity expansion tank
56 l
Overall dimensions WxDxH **
800x1007x2200 mm
Power supply (3 Phase)
400V 3~ 50Hz
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/25
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit covers a very wide operating temperature range from 65 to 300 °C, so it suits high-temperature processes rather than sub-ambient work.
Defines the process window the circulator can serve and rules it out for cooling below 65 °C.
Source
“Operating temperature range
65... 300 °C”
Performance
Heating power is 96 kW and cooling power at 300 °C is 30 kW, so the heat exchanger and facility power must be sized for these loads.
Determines electrical supply sizing and the achievable heat-up/cool-down rates for the process.
Source
“Heating power
96 kW”
Installation
The unit requires a 400 V three-phase 50 Hz supply, so the site must provide a suitable three-phase feed.
Dictates the electrical infrastructure and rules out single-phase installation.
Source
“Power supply (3 Phase)
400V 3~ 50Hz”
Installation
Three-phase devices drawing less than 63 A are delivered with cable but without a plug, so the customer must provide and fit the plug.
Affects site preparation and the electrical bill of materials for commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances are coupled: ±5 % voltage only with ±2 % frequency, and opposite-sign deviations are not allowed.
Prevents selecting a supply point whose combined voltage/frequency deviation would violate the permitted envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for high leakage.
Impacts earth connection design and choice of residual current protection.
Source
“Attention: leakage current > 3,5mA”
Conformité
The equipment is EMC class A group 1, so it is intended for industrial environments and may cause interference in residential areas.
Constrains where the unit may be installed and what EMC measures are needed.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The enclosure is IP20, so it must be installed in a dry, protected location away from water and dust ingress.
Limits the permissible installation environment and enclosure requirements.
Source
“Degree of Protection
IP20”
Operation
Ambient temperature must stay between 5 and 40 °C, so the room climate must be controlled for reliable operation.
Defines the site climate envelope and cooling/ventilation needs of the installation room.
Source
“min. ambient temperature
5 °C”
Installation
The unit is 800 x 1007 x 2200 mm, so space and access around it must be planned accordingly.
Needed for layout, transport route and maintenance access planning.
Source
“Overall dimensions WxDxH **
800x1007x2200 mm”
Selection
The pump delivers up to 200 l/min at 5.5 bar, so external circuit pressure drop must be checked against this curve.
Determines whether the pump can serve the external loop and what pressure losses are acceptable.
Source
“max. delivery
200 l/min”
Selection
Maximum permissible kinematic viscosity is 50 mm²/s, so the thermofluid must be chosen within this limit at operating temperature.
Constrains thermofluid selection and pump loading.
Source
“max. permissible kin. viscosity
50 mm²/s”
Maintenance
The expansion tank holds 56 l and is not thermoregulated, so fluid expansion volume and venting must be planned.
Affects fluid fill volume, expansion allowance and system venting design.
Source
“Filling capacity expansion tank
56 l”
Safety
The unit has adjustable overtemperature protection per DIN 12876, so the safety limit must be set for the application.
Requires commissioning step to set the overtemperature cut-out for safe operation.
Source
“With adjustable overtemperature protection according to DIN 12876.”