Improving Food Safety Analysis with LC Q TOF Feb2020

Technical Reference Summary

Covered Models & Document Scope

Ebook LCGC patrocinado (febrero de 2020) que abarca flujos de trabajo de análisis de seguridad alimentaria con el Agilent 6546 LC/Q-TOF, el Agilent 1290 Infinity II LC, MassHunter Quant 10.1 y LC/Q-TOF Screener Tool; incluye análisis de pesticidas en matrices alimentarias complejas, cribado no dirigido de adulterantes/contaminantes y pruebas de autenticidad alimentaria con MassHunter Classifier. No se indican códigos de pedido de instrumentos ni variantes de tamaño.

Technical Data

Parameter Value Condition
Mass resolution type high mass resolution p.3 ↗ Agilent 6546 LC/Q-TOF
Mass accuracy high mass accuracy p.3 ↗ Agilent 6546 LC/Q-TOF
Mass accuracy requirement for identification less than 5 ppm p.4 ↗ SANTE/12682/2019, two ions
Minimum number of ions for identification two ions with at least one fragment ion p.4 ↗ SANTE/12682/2019
Minimum acquisition rate for quantitation 2 Hz p.5 ↗ 12 points over 6 s chromatographic peak
Data acquisition speed 6 Hz p.7 ↗ All Ions acquisition, strawberry method
Chromatographic peak width about six s p.5 ↗ typical peak
Dynamic range channels two channels (low-gain and high-gain), dual-gain 10-GHz spectrum p.5 ↗ Agilent 6546 LC/Q-TOF
LC back pressure rating 1,300 bar p.3 ↗ Agilent 1290 Infinity II LC
Detector data rate 240 Hz p.3 ↗ VWD and DAD
Autosampler carryover less than 9 ppm p.3 ↗ Agilent 1290 Infinity II LC
Autosampler sample capacity 6,144 samples in 16x384 well plates p.3 ↗ Agilent 1290 Infinity II LC
SureMass data conversion time 3–5 minutes per sample p.6 ↗ post-acquisition script
Number of pesticides quantified 180 p.6 ↗ proof of concept, four matrices
Number of pesticides screened over 200 p.6 ↗ proof of concept, four matrices
Mass error achieved less than 2 ppm p.8 ↗ almost all analytes at two concentrations
Column length 150 mm p.7 ↗ C18 column with guard column
Extraction solvent volume 10 mL p.7 ↗ acetonitrile, strawberry QuEChERS

Operating Limits

  • minimum mass accuracy for identification: less than 5 ppm p.4 ↗
  • minimum acquisition rate for quantitation: 2 Hz p.5 ↗
  • LC maximum back pressure: 1,300 bar p.3 ↗
  • autosampler carryover: less than 9 ppm p.3 ↗