Cajas de guantes con filtro HEPA Precise™ de Labconco, cajas de pesaje XPert®, cajas de guantes de atmósfera controlada Precise™, extractores portátiles FilterMate® y accesorios; información de pedido páginas 8, 9, 12, 17-20

Technical Data

Parameter Value Condition
Air volume changes per minute 1.0 p.8 ↗ Precise HEPA-Filtered / XPert Weigh, Min. Dial
Air volume changes per minute 7.7 p.8 ↗ Precise HEPA-Filtered / XPert Weigh, Max. Dial
Exhaust volume 13 CFM p.8 ↗ Precise HEPA-Filtered / XPert Weigh, Min. Dial
Exhaust volume 100 CFM p.8 ↗ Precise HEPA-Filtered / XPert Weigh, Max. Dial
Noise pressure 44 dB(A) p.8 ↗ Precise HEPA-Filtered / XPert Weigh, Min. Dial
Noise pressure 71 dB(A) p.8 ↗ Precise HEPA-Filtered / XPert Weigh, Max. Dial
Static pressure in glove box 0.14 in. of H2O p.8 ↗ Precise HEPA-Filtered / XPert Weigh, Min. Dial
Static pressure in glove box 0.91 in. of H2O p.8 ↗ Precise HEPA-Filtered / XPert Weigh, Max. Dial
Total system static pressure 0.34 in. of H2O p.8 ↗ Precise HEPA-Filtered / XPert Weigh, Min. Dial
Total system static pressure 2.42 in. of H2O p.8 ↗ Precise HEPA-Filtered / XPert Weigh, Max. Dial
Fluorescent lamp power 22 watts p.8 ↗ Precise HEPA-Filtered Glove Boxes, all models
Glove thickness 1/4 inch p.6 ↗ Neoprene gloves, polyethylene
Airflow potential 250 CFM p.13 ↗ FilterMate 3970000, 115 volts, 60 Hz
Airflow potential 250 CFM p.13 ↗ FilterMate 3970020, 230 volts, 50 Hz
Airflow potential 200 CFM p.13 ↗ FilterMate 3970003, 115 volts, 60 Hz
Electrical requirements 115 volts, 60 Hz p.13 ↗ FilterMate 3970000
Electrical requirements 230 volts, 50 Hz p.13 ↗ FilterMate 3970020
Overall dimensions 14.0" w x 23.0" d x 22.7" h p.13 ↗ FilterMate 3970000
Shipping weight 50/22.7 lbs./kg p.13 ↗ FilterMate 3970000
Displacement at 60 Hz (50 Hz) 117 (98) p.18 ↗ Vacuum pump 1472100, 115 volts, 50/60 Hz, 4.6 A
Displacement at 60 Hz (50 Hz) 117 (98) p.18 ↗ Vacuum pump 7739402, 230 volts, 50/60 Hz, 2.4 A
Shipping weight 62 (28) lbs. (kg) p.18 ↗ Vacuum pump 1472100 / 7739402
Leak rate 1 x 10-4 ml/second p.3 ↗ Precise Controlled Atmosphere Glove Boxes, helium mass spectrometer test
Oxygen level less than 5 ppm p.16 ↗ Precise Controlled Atmosphere Glove Boxes

Operating Limits

  • Leak rate (helium mass spectrometer): no detectable leaks greater than 1 x 10-4 ml/second (8.6 ml/day) p.3 ↗
  • Oxygen level in controlled atmosphere boxes: less than 5 ppm p.16 ↗
  • Duplex receptacle maximum current: 5 amp p.8 ↗
  • Glove box atmosphere containment class: Class I per ISO 10648-2 p.3 ↗

Ordering Data

Ordering Code Part No. Variant / Code S'applique à Note
3970000 p.13 ↗ 3970000 115 volts, 60 Hz; HEPA included FilterMate Portable Exhauster, 250 CFM, room exhaust
3970020 p.13 ↗ 3970020 230 volts, 50 Hz; HEPA included FilterMate Portable Exhauster, 250 CFM, room exhaust
3970001 p.13 ↗ 3970001 115 volts, 60 Hz; Carbon required FilterMate Portable Exhauster, 250 CFM, room exhaust
3970021 p.13 ↗ 3970021 230 volts, 50 Hz; Carbon required FilterMate Portable Exhauster, 250 CFM, room exhaust
3970002 p.13 ↗ 3970002 115 volts, 60 Hz; HEPA included FilterMate Portable Exhauster, 250 CFM, 6.06" ID thimble to outside
3970022 p.13 ↗ 3970022 230 volts, 50 Hz; HEPA included FilterMate Portable Exhauster, 250 CFM, 6.06" ID thimble to outside
3970003 p.13 ↗ 3970003 115 volts, 60 Hz; HEPA included FilterMate Portable Exhauster, 200 CFM, room exhaust
3970023 p.13 ↗ 3970023 230 volts, 50 Hz; HEPA included FilterMate Portable Exhauster, 200 CFM, room exhaust
3970004 p.13 ↗ 3970004 115 volts, 60 Hz; Two carbon required FilterMate Portable Exhauster, 200 CFM, room exhaust
3970024 p.13 ↗ 3970024 230 volts, 50 Hz; Two carbon required FilterMate Portable Exhauster, 200 CFM, room exhaust