Protector pass through laboratory hoods brochure 4 55

Technical Reference Summary

Covered Models & Document Scope

Campanas de laboratorio de paso Protector® en anchos de 4', 5' y 6'; modelos sin guarniciones y con guarniciones; variantes eléctricas 100-115V 50/60 Hz 10 A y 208-230V 50/60 Hz 5 A; poste de servicio opcional trasero o lateral; ventana de visualización de vidrio opcional derecha o izquierda.

Technical Data

Parameter Value Condition
Hood width 48.0" (121.9 cm) in (cm) p.6 ↗ 4' Hood
Hood width 60.0" (152.4 cm) in (cm) p.6 ↗ 5' Hood
Hood width 72.0" (182.9 cm) in (cm) p.6 ↗ 6' Hood
Height 38.25" (97.2 cm) in (cm) p.6 ↗ 4' Hood
Height 50.25" (127.6 cm) in (cm) p.6 ↗ 5' Hood
Height 62.25" (158.1 cm) in (cm) p.6 ↗ 6' Hood
Depth 24.0" (61.0 cm) in (cm) p.6 ↗ 4' Hood
Depth 30.0" (76.2 cm) in (cm) p.6 ↗ 5' Hood
Depth 36.0" (91.4 cm) in (cm) p.6 ↗ 6' Hood
Sightline height (work surface to header panel) 37.5" in p.4 ↗
Sash opening height (with sash stop) 25" in p.4 ↗
Maximum sash opening height 28" in p.4 ↗ loading/unloading only
Airflow 775 CFM p.5 ↗ 4' Hood, s.p. 0.33
Airflow 1020 CFM p.5 ↗ 5' Hood, s.p. 0.42
Airflow 1250 CFM p.5 ↗ 6' Hood, s.p. 0.59
Airflow 590 CFM p.5 ↗ 4' Hood, s.p. 0.19
Airflow 775 CFM p.5 ↗ 5' Hood, s.p. 0.24
Airflow 950 CFM p.5 ↗ 6' Hood, s.p. 0.34
Electrical requirement 100-115V, 50/60 Hz, 10 A V, Hz, A p.6 ↗
Electrical requirement 208-230V, 50/60 Hz, 5 A V, Hz, A p.6 ↗
Lighting T8 fluorescent p.5 ↗ pre-wired
Optional viewing window size 14.9" x 23.9" in p.4 ↗ right or left mounted

Operating Limits

  • Maximum sash opening height: 28 in p.4 ↗
  • Sash stop opening height: 25 in p.4 ↗
  • Electrical supply: 100-115V or 208-230V, 50/60 Hz p.6 ↗