PHD 4 Capillary Leak
Technical Reference Summary
Covered Models & Document Scope
Agilent PHD-4 Capillary Leak, modelo n.º 9693540, accesorio de calibración para el detector de fugas Agilent PHD-4 con sonda de longitud estándar
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Part number | 9693540 p.1 ↗ | PHD-4 Capillary Leak |
| Reservoir volume | 110 cc p.1 ↗ | stainless steel reservoir |
| Pressure gauge range | 160 psig p.1 ↗ | |
| Helium concentration | of the order of 100 ppm p.1 ↗ | reservoir pressure 100 PSI, 7 bar |
| Helium recharging valve connection | Gc 1/4 p.2 ↗ | |
| O-ring | Parker 2-120 BUNA (2 pcs.) p.1 ↗ | supplied in nylon bag |
| Stabilization time | at least 5 minutes p.2 ↗ | before calibration |
| Zero reading wait time | usually about 10 sec p.2 ↗ | N2 purge until 0.0 PPM |
Operating Limits
- Pressure gauge maximum: 11 bar (160 psig) p.1 ↗
- Calibration curve validity: PHD-4 devices with standard-length probe only p.1 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| PHD-4 Capillary Leak p.1 ↗ | 9693540 | Calibration accessory for PHD-4 leak detector |