c391-e101
Technical Reference Summary
Covered Models & Document Scope
Analizador transportable de NOx-O2 NOA-7100 (documento C391-E101A). Las variantes difieren en los componentes de medición (NOx, O2 frente a NOx, NO, NO2), el rango de medición, la cantidad de muestra de gas, el separador de drenaje externo, el conjunto de caudal reducido, el conjunto de comunicación y el conjunto de válvula solenoide de conmutación NOx-NO. La unidad de pretratamiento CFP-8000 es un P/N independiente.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Measurement components | NOx, O2 p.8 ↗ | NOA-7100 |
| Measurement components | NOx, NO, NO2 (*1) p.8 ↗ | NOA-7100 variant |
| Measurement range | NOx: 0-25/50/100/250/500/1000/2500/4000 p.8 ↗ | NOA-7100 |
| Measurement range | NOx, NO, NO2: 250/500/1000/2500/4000 ppm ppm p.6 ↗ | NOA-7100 variant |
| Gas sample amount | Approx. 2 L/min p.8 ↗ | NOA-7100 |
| Gas sample amount | 100 mL/min p.8 ↗ | NOA-7100 variant |
| External drain separator | Fitted p.8 ↗ | NOA-7100 |
| External drain separator | Not fitted p.8 ↗ | NOA-7100 variant |
| Small flow rate set | Not fitted p.9 ↗ | NOA-7100 |
| Small flow rate set | Fitted p.9 ↗ | NOA-7100 variant |
| Communication set | Not fitted p.9 ↗ | NOA-7100 |
| Communication set | Fitted p.9 ↗ | NOA-7100 variant |
| NOx-NO switching solenoid valve set | Not fitted p.9 ↗ | NOA-7100 |
| NOx-NO switching solenoid valve set | Fitted p.9 ↗ | NOA-7100 variant |
| Coexisting gas CO | - p.9 ↗ | NOx, NO, NO2 measurement |
| Coexisting gas CO | Less than 1 vol% p.9 ↗ | O2 measurement |
| Coexisting gas CO2 | 30 vol% or less vol% p.9 ↗ | NOx, NO, NO2 measurement |
| Coexisting gas CO2 | Less than 30 vol% p.9 ↗ | O2 measurement |
| Coexisting gas Hydrocarbon | - p.9 ↗ | NOx, NO, NO2 measurement |
| Coexisting gas Hydrocarbon | Less than 1 vol% p.9 ↗ | O2 measurement |
| Coexisting gas N2O | - p.9 ↗ | NOx, NO, NO2 measurement |
| Coexisting gas N2O | Must not be contained p.9 ↗ | O2 measurement |
| Coexisting gas SO3 | 50 ppm or less ppm p.9 ↗ | NOx, NO, NO2 measurement |
| Coexisting gas SO3 | Must not be contained p.9 ↗ | O2 measurement |
| Coexisting gas Other flammable gases | Must not be contained p.9 ↗ | NOx, NO, NO2 measurement |
Operating Limits
- CO2 coexisting gas (NOx, NO, NO2 measurement): 30 vol% or less p.9 ↗
- CO2 coexisting gas (O2 measurement): Less than 30 vol% p.9 ↗
- CO coexisting gas (O2 measurement): Less than 1 vol% p.9 ↗
- Hydrocarbon coexisting gas (O2 measurement): Less than 1 vol% p.9 ↗
- SO3 coexisting gas (NOx, NO, NO2 measurement): 50 ppm or less p.9 ↗
- N2O coexisting gas (O2 measurement): Must not be contained p.9 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| CFP-8000 p.10 ↗ | CFP-8000 | pretreatment unit |