Optimizing Microplastic Characterization by LDIR: Automated versus Manual Workflows
Technical Reference Summary
Covered Models & Document Scope
Resumen técnico del sistema de imágenes químicas Agilent 8700 LDIR con el software de control Clarity; abarca el flujo de trabajo de análisis de partículas (automatizado) con un intervalo de tamaños predeterminado de 20 a 500 µm y el flujo de trabajo de análisis manual; evaluado con microesferas de látex de poliestireno trazables según NIST de 2, 5 y 10 µm sobre portaobjetos low-E (número de pieza de Agilent M7300-68010) y filtros de poliéster recubiertos de Al (número de pieza de Agilent M7300-68011)
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Default particle size range (Particle Analysis workflow) | 20 to 500 µm p.1 ↗ | adjustable based on user methodology |
| Minimum detectable particle size, automated mode | 5 µm p.7 ↗ | 5 µm PS beads, low-E slides and filters |
| IR wavenumber used for imaging | 1,442 cm–1 p.3 ↗ | |
| Low-E IR reflective slide dimensions | 25 × 75 mm p.2 ↗ | Agilent part number M7300-68010 |
| Al-coated polyester filter diameter | 25 mm p.2 ↗ | Agilent part number M7300-68011 |
| Al-coated polyester filter pore size | 0.8 µm p.2 ↗ | Agilent part number M7300-68011 |
| Al-coated polyester filter coating | 100/0 nm p.2 ↗ | Agilent part number M7300-68011 |
| PS latex bead sizes evaluated | 2, 5, and 10 µm p.1 ↗ | NIST-traceable |
Operating Limits
- Default particle size range (Particle Analysis workflow): 20 to 500 µm p.1 ↗
- Minimum detectable particle size, automated mode, low-E slides: 2 µm p.7 ↗
- Minimum detectable particle size, automated mode, low-E slides and filters: 5 µm p.7 ↗
- Minimum detectable particle size, automated mode, low-E slides and filters: 10 µm p.7 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | S'applique à | Note |
|---|---|---|---|---|
| Low-E infrared (IR) reflective slide p.2 ↗ | M7300-68010 | 25 × 75 mm | Substrate used in the study | |
| Al-coated polyester filter p.2 ↗ | M7300-68011 | 25 mm, 100/0 nm coating, 0.8 µm pore size | Substrate used in the study |